15
sc100 Operation
4.2 Sensor Setup
When a sensor is initially installed, the serial number of the sensor will be displayed as the
sensor name. To change the sensor name refer to the following instructions:
4.2.2 Sensor Data Log
The data logs store the measurement data at selected intervals. The data logs are stored
in a packed binary format and the event logs are stored in a CSV format. The logs can be
downloaded through the digital network port, service port, or the IrDA port. DataCom is
needed for downloading logs to a computer.
4.2.1 Changing the Sensor Name
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
SENSOR SETUP
3
Highlight the appropriate sensor if more than one sensor is attached.
4
CONFIGURE
5
EDIT NAME
6
Select character to edit (shown in brackets).
—
Scroll to appropriate character.
4.2.2.1 Sensor Data Logging
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
SENSOR SETUP
3
CONFIGURE
2
LOG SETUP
4
SENSOR INTERVAL
2
Scroll through the displayed options.
4
TEMP INTERVAL
2
Scroll through the displayed options.
5
SET INTERVAL
客服:400 688 9809
Summary of Contents for 5740sc
Page 1: ...Visit http www hach com 客服 400 688 9809 ...
Page 6: ...Visit us at www hach com 客服 400 688 9809 ...
Page 12: ...12 Installation Figure 4 Sensor Installation Examples 客服 400 688 9809 ...
Page 22: ...Visit us at www hach com 客服 400 688 9809 ...
Page 34: ...Visit us at www hach com 客服 400 688 9809 ...
Page 38: ...Visit us at www hach com 客服 400 688 9809 ...
Page 40: ...Visit us at www hach com 客服 400 688 9809 ...
Page 44: ...Visit us at www hach com 客服 400 688 9809 ...