Get a sample measurement
Use this option to get sample measurements with one of the installed applications.
1.
Make sure that the glass beaker, the sensor holder and all related parts are clean and dry. Put a
magnetic stir bar into the beaker.
2.
Attach the beaker to the sensor holder. Make sure that the beaker fits tightly into the sensor
holder.
3.
Read the related "Application Note" from the USB applications key for more instructions.
4.
Fill the beaker with the applicable amount of solvent. Refer to
on page 18.
Make sure that there is no liquid spill. Make sure that the initial level is enough so that the sensor
is correctly installed in the sample. Do not put too much solvent in the beaker.
5.
From the main menu, select the measurement application, then push
Start
. Application
information shows on the display.
6.
If necessary, select an icon for more information or to change some data.
7.
Make sure that the icon at the bottom of the display
is highlighted. Do the instructions that
show on the display adjacent to this icon. Make sure that the tubes and sensor are correctly
aligned.
8.
Push
Start
to start the cell conditioning. Measurement data shows on the display.
9.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: Changing the default stirring speed can effect the measurement result.
10.
If the options are available during the measurement procedure, push
Skip
to ignore the current
step or push
Stop
to abort the measurement.
11.
Weigh the sample in its container and write down the result.
12.
When the cell conditioning is complete add the sample. Remove the sample stopper and put the
correct amount of sample into the beaker with an applicable utensil. Make sure that all of the
sample is put into the center of the beaker and that none of it spills onto the walls of the beaker.
13.
Put the sample stopper back into the sensor holder. Make sure that the sample stopper fits tightly
into the sensor holder.
14.
Weigh the empty sample container and subtract this value from the weight of the sample in its
container (refer to step
). The result is the exact weight of the sample analyzed. Write down
this value.
15.
Push
Start
to start the moisture content analysis. Measurement data shows on the display.
Note: The moisture content analysis will start automatically after the addition of sample if the Injection
Autodetect parameter is set to Yes.
16.
If the default stirring speed needs to be adjusted, push the up and down arrow keys to increase or
decrease the speed.
Note: Changing the default stirring speed can effect the measurement result.
17.
If the options are available during the measurement procedure, push
Skip
to ignore the current
step or push
Stop
to abort the measurement.
18.
When the measurement is complete, enter the exact weight of the sample analyzed (refer to step
). Push the arrow keys to see the different measurement views.
19.
Push
Next
for the options that follow:
Option
Description
Replicate sample
Use this option to start the same titration on the same sample. This is used to study the
repeatability by successively analyzing several parts of the same sample. At the end of
each measurement, a window shows the average value, the standard deviation and the
relative standard deviation.
English
17
Summary of Contents for TitraLab KF1121
Page 21: ...English 21...
Page 130: ...130 Portugu s...
Page 151: ...e tina 151...
Page 172: ...172 Dansk...
Page 237: ...Svenska 237...
Page 262: ...1 2 15 1 2 3 4 5 1 262...
Page 263: ...1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 2 USB USB 263...
Page 264: ...USB USB 2 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 3 1 3 3x 5 2 4 264...
Page 265: ...4 1 KF 6 2 KF Teflon 7 3 10x 8 4 USB 9 3 x GL45 5 10 3x 1 15 6 in 1 2 1 265...
Page 266: ...GFCI GFI 1 2 2 264 3 180 266...
Page 267: ...1 263 267...
Page 268: ...268...
Page 269: ...269...
Page 270: ...MSDS SDS 270...
Page 271: ...278 1 2 271...
Page 272: ...5 5 1 4 2 5 3 2 272...
Page 273: ...MSDS SDS 1 Settings 2 Select 273 273 3 Back 1 Settings 2 Applications Edit 273...
Page 275: ...MSDS SDS a 1 2 3 USB 4 276 5 Start 6 7 8 Start 9 10 Skip Stop 11 12 13 275...
Page 277: ...1 Data log 2 Select USB 3 1 Calibration 2 3 Start 4 5 6 Continue 7 Yes No 8 Exit 277...
Page 278: ...274 300 C 572 F 4 Maintenance 1 Pump activation 278...
Page 279: ...Syringe replacement 266 1 Pump activation 2 Toggle 279...
Page 280: ...1 Select Reset 280...
Page 281: ...Edit 281...
Page 322: ...322 Rom n...
Page 343: ...T rk e 343...
Page 364: ...364 Slovensk jazyk...
Page 385: ...Slovenski 385...
Page 406: ...406 Hrvatski...
Page 410: ...1 2 15 FCC 1 2 3 4 5 1 410...
Page 411: ...1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 2 USB USB 411...
Page 412: ...USB USB 2 1 24 V 4 7 USB 2 1 5 8 Ethernet 3 2 6 3 1 3 x3 5 2 4 412...
Page 413: ...4 1 KF 6 2 KF Teflon 7 3 x10 8 4 USB 9 3 x GL45 5 10 3x 1 15 cm 1 2 1 413...
Page 414: ...AC GFCI GFI 1 2 2 412 3 180 414...
Page 415: ...1 411 415...
Page 416: ...416...
Page 417: ...417...
Page 418: ...MSDS SDS 418...
Page 419: ...426 1 2 419...
Page 420: ...5 5 1 4 2 5 3 2 420...
Page 423: ...MSDS SDS 1 2 3 USB 4 424 5 Start 6 7 8 Start 9 10 Skip Stop 11 423...
Page 426: ...3 Start 4 5 6 Continue 7 Yes No 8 Exit 422 300 C 4 426...
Page 428: ...428...
Page 431: ......