
25
10.
Press
GREEN
key. Enter "01.50" by using
BLUE
key to change value and
GREEN
key to move to the next space. Press
GREEN
key until "A1" is
displayed.
11.
Press
GREEN
key. Enter "0.728" by using
BLUE
key to change value and
GREEN
key to move to the next space. Press
GREEN
key until "Add" is
displayed.
12.
Press
MENU
until "Edit" is displayed. Press
GREEN
key until „S0“ is displayed.
13.
Press
GREEN
key. Enter "00.05" by using
BLUE
key to change value and
GREEN
key to move to the next space. Press
GREEN
key until "A0" is
displayed.
14.
Press
GREEN
key. Enter "0.090" by using
BLUE
key to change value and
GREEN
key to move to the next space. Press
GREEN
key until „S1“ is
displayed.
15.
Press
MENU
until "Edit" is displayed to check the entered data. Press
GREEN
key. „RES“ is displayed.
16.
Press
BLUE
key to scroll through the entries. Compare the entries with the data
shown in the table below.
17.
After comparing the entries press
MENU
two times to enter the measuring
mode. See
4.3 Evaluate LANGE Phosphate-Phosphorus/ Phosphate Cuvette
.
Note: In case of any mistake retrieve the factory calibration first. See
. After retrieving the factory calibration, start the programming procedure
again. If the instrument is shut-off or loses power during data entry, all edits will be lost.
Automatic shut-off is 60 minutes.
3.3.3 LCK 350 Phosphate-Phosphorus PO
4
-P
1.
Press
POWER
to switch on the instrument.
2.
Press
MENU
. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2).
3.
Select measuring range 1 or 2 for LCK 350 (2-20 mg/l PO
4
-P).
Note: Press
GREEN
key to toggle between measuring ranges.
4.
Press
MENU
for about five seconds until "USER" followed by "CAL" is
displayed.
5.
Press
BLUE
key until "Edit" is displayed. Press
GREEN
key. " RES" is
displayed.
6.
Press
GREEN
key. "-.---" is displayed.
Table 4 Calibration data of LCK 349
RES
--.--
S0
0.05
A0
0.090
S1
1.50
A1
0.728
Summary of Contents for POCKET COLORIMETER II Phosphate LCK 348
Page 2: ...2 ...
Page 4: ...4 ...
Page 6: ...6 ...
Page 8: ...8 ...
Page 10: ...10 ...
Page 12: ...12 ...
Page 14: ...14 ...
Page 16: ...16 ...
Page 18: ...18 ...
Page 20: ...20 ...
Page 28: ...28 ...
Page 32: ...32 ...
Page 34: ...34 ...
Page 36: ...36 ...
Page 38: ...38 ...
Page 46: ...46 ...
Page 50: ...50 ...
Page 52: ...52 ...
Page 54: ...54 ...
Page 56: ...56 ...
Page 64: ...64 ...
Page 68: ...68 ...
Page 70: ...70 ...
Page 72: ...72 ...
Page 74: ...74 ...
Page 82: ...82 ...
Page 86: ...86 ...
Page 88: ...88 ...
Page 90: ...90 ...
Page 92: ...92 ...
Page 100: ...100 ...
Page 104: ...104 ...
Page 106: ...106 ...
Page 108: ...108 ...
Page 110: ...110 ...
Page 118: ...118 ...
Page 122: ...122 ...
Page 124: ...124 ...
Page 126: ...126 ...
Page 128: ...128 ...
Page 138: ...138 ...
Page 140: ...140 ...