Page 8 / 64
OCTOPUS 300 Service Manual
3.3
Installation
3.3.1
Instrument Table
The instrument table is delivered in a separate package. Utilize the instructions included with the table to put the table together and
take care to select the correct voltage before connecting the power cord.
Figure 6
6
Voltage selection switch on the electronics box
7
Power connector for attaching the OCTOPUS 300
WARNING:
Be sure to set the correct voltage before
inserting the power cord connector.
Figure 7
Power connector socket with fuse holder on the electronics box
3.3.2
OCTOPUS 300
Handle the instrument using both lower housing halves to lift it out of the packing. Two ribbed grips hinder sideways slipping.
Since the OCTOPUS 300 works without a cupola, a fully darkened room is not required. In order, though, to make the examination
conditions pleasant for the patient and for obtaining reliable results, the instrument is to be placed in the room so that no direct light
falls on the instrument or the patient.
The positioning between the patient and the operator or the operating panel can be so chosen that the room conditions are optimally
used.
6
7
Do not use the forehead rest as a carrying
handle during unpacking or transportation!
Stirnstütze beim Auspacken oder
Transportieren nicht als Traggriff benutzen!
Ne pas utiliser l'appui-tête comme poignée
pendant le déballage ou pour transporter!