Tradition and Innovation
Page 10 / 64
OCTOPUS 300 Service Manual
4
S
OFTWARE
4.1
Installation, Exchange, Upgrade
For application software installation and upgrade refer to Chapter 7.14.1
'SIM Flash Module, Application Software (1802222).
4.2
Releasing Program Options
Some of the functions of the OCTOPUS 300 are offered as options and are not accessible in the basic version of the software.
Basically all the functions are present in the software, but some can only be made accessible with a code. This code is delivered by
HAAG STREIT and is then entered by the customer or the service technician into the instrument. Information about such additional
functions can be obtained directly from HAAG STREIT or from your representative.
4.2.1
Procedure
Dongle codes
Read out software master data
Actual dongle code
Serial no.
Software version
Enter new dongle code. The keyboard is
only released, when the serial number of
the instrument is stored in the memory.
To print out the complete instrument
information see Chapter 8.8
Procedure to make additional functions accessible in the perimeter.
IMPORTANT:
The 'Dongle code' function is only used, when additional program options have been purchased. To avoid
free access and faulty manipulations the function is protected. The 'Dongle' button is only released after the
two headrest sensors and the 'Dongle' button are pressed at the same time. As from Serial No. 1031 the
head rest sensors have been replaced with non-contact head sensors. In this case hold your hand (close) in
front of the head rest and simultaneously press the ‘Dongle’ button.
Customer
Sends the serial no., software version and the complete dongle code to HAAG STREIT or the
representative and orders the desired supplementary function(s).
HAAG STREIT AG
Based on the serial no. and the present and ordered option(s), the new dongle code is computed
and given to the customer.
Customer
Enters the complete new dongle code into the OCTOPUS 300.
Open dongle code tab.
Click on 'Dongle code' entry field (numerical keyboard will be displayed).
Enter complete dongle code into the three entry fields.
Close keyboard with 'Esc'.
Save new dongle code by clicking on
Before the instrument saves the setting, the parameters are verified by the system. A faulty
dongle code will not be saved. The message
'Dongle Code not accepted'
is displayed on the
screen. Delete wrong code and enter it the correct way.
OCTOPUS 300
The new additional functions are accessed via the graphical user interface.