INSTALLATION GUIDE
LENSTAR LS 900
AND
LENSTAR LS 900 APS
Edition 3.10 / 2016-02
Page 1: ...INSTALLATION GUIDE INSTALLATION GUIDE LENSTAR LS 900 AND LENSTAR LS 900 APS Edition 3 10 2016 02 ...
Page 2: ...ortant information Operation of the appliance for performance of an examination as well as the description of the EyeSuite software are described in a separate operating manual and can be called up in the EyeSuite software using the F1 key Instructions regarding the function check can be found in the Instructions for use 7220055 EyeSuite Setup manual 7220060 regarding the installation of EyeSuite ...
Page 3: ...3 Contents 1 Safety Instructions 4 1 1 Safety Instructions in text 4 2 Installation 4 2 1 Safe system configuration according to EN 60601 1 4 2 2 Sample configuration according to EN 60601 1 5 2 3 Mounting the roller rails slide plate and cable connection 6 2 4 Instrument table option assembly 13 List of modifications 14 ...
Page 4: ...d patients WARNING These signs must be strictly adhered to to ensure safe operation of the instrument and to avoid endangering users and patients NOTE Important notes Please read carefully 2 Installation 2 1 Safe system configuration according to EN 60601 1 FORBIDDEN It is not permitted to make modifications to the measuring device WARNING The device must only be installed on a stable horizontal s...
Page 5: ...S Doc No 7220553 Edition 3 10 2016 02 5 2 2 Sample configuration according to EN 60601 1 1 Power Supply LENSTAR 1020392 LENSTAR APS 1022106 2 Laptop PC 3 Instrument Table 4 Power Supply Laptop PC 5 Insulation transformer 6 Printer med approved EN 60601 1 USB 2 0 or higher 1 2 3 4 6 5 ...
Page 6: ...urface NOTE Function check and right left recognition test must be performed after every new installation changes on the rail changes of side identification sticker 2 3 1 Roller rails slide plate for LENSTAR LS 900 not for LS 900 APS The roller rails slide plate must be positioned in such a way that when the device is centred the patient s eye will be at a distance of 78 mm from the front of the d...
Page 7: ... need either the HAAG STREIT instrument table HSM 901 option or an aperture in the table top used Remove cable cover 1 Expose USB and supply cables 2 Insert the supply cable through the opening beneath the cross slide 3 until the cable emerges from the opening at the top at the front Insert the USB cable through the same opening 3 on the cross slide Insert both cables in the appropriate box and fi...
Page 8: ...es left or right aperture on table With a knife cut carefully into the snap off point provided on the left or right of the cable cover 6 Carefully use fingers to snap off the cut part 7 Use a spiral wrap or the black tube 1003686 to join the USB and supply cable together 8 Insert both cables in the appropriate box and fix with the tension compensator Mount cable cover Mount the rail covers 1021914...
Page 9: ...21556 1x Magnetic slide plate 1022332 1x Side identification sticker 1022151 NOTE Clean the table surface before positioning the side identification sticker 1022151 Make sure that no oil and grease residue remains on the surface WARNING The aperture of the cables has to be cut free form the side identification sticker with a sharp knife Ensure that there are not white or reflecting edges when doin...
Page 10: ... the left side seen form the joystick Expose USB and supply cables table aperture 9 The free cable length should not exceed 150mm Connect the supply and USB cable to the LENSTAR as seen in image 10 Secure the connectors and cables as shown in image 11 using standard cable binders Make sure that the flaps are not mounted downwards as they may drag on the table otherwise Place the LENSTAR on the rai...
Page 11: ...e for LENSTAR LS 900 APS Remove the rail covers Place the LENSTAR APS on the right side seen form the joystick Remove the cable channel cover on the right side of the LENSTAR using a 1 5mm hex key 12 Snap off the cable channel cover piece and smoothen the edge with a sharp knife 13 Pull the two cables through the black tube 14 1003686 Figure 12 Figure 13 Figure 14 ...
Page 12: ...to the LENSTAR Secure the connectors and cables as shown in image 15 using standard cable binders Make sure that the flaps are not mounted downwards as they may drag on the table otherwise Mount the cable channel cover using a 1 5mm hex key 16 Mount the rail covers Pull the black tube through the cable guide 17 Check that the motion of the LENSTAR is not hindered by the cabling 18 Figure 15 Figure...
Page 13: ...instructions for use enclosed with it The LENSTAR LS 900 APS Sets with table HSM 901 Workstation includes two spacers 1022335 19 The Workstation table top is 4mm thicker than the Standard table top To compensate this differences this spacers are mounted Figure 19 To mount the spacers remove the upper part of the headrest while using a 2 0mm hex key 20 insert the spacer on the right and left side 2...
Page 14: ...3 10 2016 02 List of modifications Edition Date Changes 3 10 08 02 2016 Complement of drawing Aperture for cables in chapter 2 3 4 3 00 05 06 2015 Add quantity information on components of the table fitting set 1022328 in chapter 2 3 4 2 00 08 05 2015 Move and change of chapter 2 4 1 00 01 05 2015 First release ...
Page 15: ......
Page 16: ...For further questions please contact the HAAG STREIT Customer Support at service haag streit com ...