Check the drum position 0
In working position, the feeler arm is swivelled and the measuring drum is set to 0.
The arm carrying the measuring prism must swing freely between the limit stops on
contact.
Check the drum position 2
•
The control weight is to be used for this inspection. The weight rod is engraved
with five rings. The middle ring corresponds to a scale value of 0, the rings im
‑
mediately to its left and right correspond to a value of 2, and both outside scale
rings have a value of 6.
•
A mark corresponding to the measuring drum setting 2 is set exactly to the index
mark on the weight holder, whereupon the weight is fitted on the tonometer axis
so that the longer part points in the direction of the patient. The feeler arm must
move towards the patient.
6
2 0
6
2
•
Check position 1.95
Move calibration mark 2 on the measuring drum downwards, opposite the fixed in
‑
dex, by a width of one marking. With light pressure to the feeler arm in the direction
of free movement, the feeler arm must start to move independently towards the limit
stop in the direction of the examiner.
•
Check position 2.05
Move calibration mark 2 on the measuring drum upwards, opposite the fixed index,
by a width of one marking. The feeler arm must start to move into direction of the
stop‑piece on the patient's side.
1.95
2.05