Example 1: Parameter setting using all correction points
Point Number
Actual Velocity (Correction) Point
Expect Velocity at this Point
Correction Value Action Range
1
0.100m/s
0.110m/s
Zero velocity -0.100m/s
2
0.150m/s
0.160m/s
0.100m/s - 0.150m/s
3
0.200m/s
0.220m/s
0.150m/s - 0.200m/s
4
0.250m/s
0.270m/s
0.200m/s - 0.250m/s
End
0.300m/s
-
0.250m/s - 0.300m/s
Example 2: Parameter setting using partial correction points
Point Number
Actual Velocity (Correction) Point
Expect Velocity at this Point
Correction Value Action Range
1
0.100m/s
0.110m/s
Zero velocity -0.100m/s
2
0.150m/s
0.160m/s
0.100m/s - 0.150m/s
3
0.161m/s
0.161m/s
No correction
4
0.162m/s
0.162m/s
No correction
End
0.163m/s
-
No correction
NOTE: When using non-linear correction, users need to set all correction points and correction numbers according to
the setting principle.
If no set correction point or correction number provided, the screen will prompt “BUG”, and the non-linear function will
not work.