10.
programming
To proceed with programming, first locate the DIP switches on the back side
of the Digital display. Refer to the circled area in Fig.3.1.
Important Note:
It is important to select the correct DIP switch position
based on
BOTH
the model (first column) and the pipe size (second column)
shown in the table below as multiple AcuFlow models exist for the same pipe
diameter.
AcuFlow Model
Pipe Size
DIP Switch Position
1
2
3
4
AF-2; AF-M-DN50
2” and DN50
ON
OFF
OFF
OFF
AF-C-S
2”
ON
ON
OFF
OFF
AF-25; AF-M-DN65
2.5” and DN65
ON
ON
ON
OFF
AF-3; AF-M-DN80
3” and DN80
OFF
OFF
ON
OFF
AF-4; AF-M-DN100
4” and DN100
OFF
OFF
OFF
ON
AF-6; AF-M-DN150
6” and DN150
OFF
ON
OFF
ON
AF-8; AF-M-DN200
8” and DN200
OFF
ON
ON
OFF
AF-15; AF-M-DN40
1.5” and DN40
OFF
ON
ON
ON
AF-15-U
1.5”
ON
ON
OFF
ON
AF-2-U
2”
OFF
OFF
ON
ON
www.h2flow.net
•
Tel: 888-635-0296 (Toll Free)
OR
(+1) 419-841-7774 (International)
11
Fig 3.2 (left)
The
default setting for
the DIP switches (all
switches set to the
OFF position).
Fig 3.3 (left)
DIP
switch number 1 set
to the ON position.
Fig 3.1