page 37 of 90
cos105000_200-300_75-125, rev. 1.03
© 1988 - 2019 h/p/cosmos sports & medical gmbh
If the heart rate signal totally fails, an acoustic warning signal occurs and the heart rate display shows no value any more.
Furthermore, the device reduces speed and elevation to 0 within one minute.
For possibilities to interfere with an automatic program, see “interfere with an automatic program”.
The treadmill adjusts speed and elevation according to following matrices.
Deviation from lower limit
Speed (km/h)
Elevation (%)
Reaction time (s)
< 5 heart beats
0.2
0.1
25
6 … 15
0.4
0.2
25
16 … 30
0.6
0.4
25
31 … 50
0.8
0.8
20
> 50 heart beats
1.0
1.0
20
Deviation from upper limit
Speed (km/h)
Elevation (%)
Reaction time (s)
< 5 heart beats
0.3
0.3
12
6 … 15
0.8
0.8
12
16 … 30
1.0
1.0
10
31 … 50
1.5
1.2
8
> 50 heart beats
2.0
1.6
7
It is possible to read out the data of the whole treatment via RS232 interface.
The minimum resolution is 1/s.
The data do not remain through restart or in case of power failure.