6.
Set emergency stop position
When subject stands upright, pull knob and move
slide down to one step before lowest possible
position.
Now the subjects’ vertical movement is limited.
Warning!
The pull knob and slide in order to adjust the
position of the fall stop cannot be re-adjusted in
case the subject is unconscious.
Thus a lowering of a subject with the help of the air
pressure control valve may not work until a subject
can reach a wheelchair position.
7.
Set unweighting force
Adjust the unweighting force by turning the control
knob until the desired value is reached by the
pointer.
Start exercise.
Warning!
Keep subject under permanent supervision.
8.
Set back emergency stop position
Instruct subject to stand on the feet and hold tight
to the treadmill handrails.
Pull knob und push slide upwards completely.
Now the subjects’ vertical movement is free.
9.
Lower subject (back to wheelchair)
After exercise, turn control knob counter-clockwise
carefully to reduce unweighting completely.
If possible, ask the subject to support the process.