ERGO LIFT 650
13
Functional test of the limit sensor:
-
Press the "Up" or "
↑
" button until a double beep is heard and the lift stops automatically.
The distance from the ground to the crossbar should be between 55.5cm and 56.5cm. If the reached height
is lower or higher, an adjustment of the sensor rod must be carried out. If you find that the motor does not
stop automatically when the lift no longer rises, the sensor rod must be adjusted as well. After adjustment,
check that the table is within the tolerances shown in the diagram above. If you still cannot find the right
position, check the connections or the status of the sensor cable from the sensor to the control board. When
working on the electrical circuit, always unplug the power cord and switch off the hydraulic unit.
By adjusting the length of the rod you will be able to find the right setting.
Checking the oil level:
The ERGO LIFT is supplied with sufficient oil, there is no need to add oil during start-up.
The tank can hold 4L of oil. A dipstick is used to check that the oil level is sufficient. Simply check that the
level is higher than the indicator on the cap (minimum level) and do not exceed this level by more than 1cm
to 2cm. Use chemical-resistant nitrile rubber protective gloves and goggles when leveling the oil.
Unless the oil is used very intensively and there is a colour change in the oil, there is no indication to replace
the oil every year. We recommend replacing the oil every 5 years at the same time as the hose.
Use high performance hydraulic oil with a viscosity of 32CST (reference 062160 / type IKV-TRIBOLINE HLQ
HM-32).
+/-
Limit switch adjustment
STOP
Going
Down
55,5 / 56,5cm
Limit switch sensor
Flat table
Drain plug
ok
1- Slightly loosen the screw (1).
2- Adjust the rod +/- (2)
3- Tighten the screw (1)
4- Check the height: 55.5 <H<56.5
(1)
(2)