Pulson 200
8
1.2.2
Electrical safety
1.2.3
Prevention of explosion
1.2.4
Electro Magnetic Compatibility
1.2.5
Ultrasound therapy
•
Only use the equipment in an area with facilities that
meet the applicable legal regulations.
•
Connect the equipment to an outlet with a protective
earth terminal. The outlet must meet the locally
applicable requirements for medical areas.
•
Do not use the equipment in an area where combustible
gases or vapours are present.
•
Switch off the equipment when it is not used.
•
Medical electrical equipment requires special precautions
for Electro Magnetic Compatibility (EMC). Follow the
instructions for the installation of the equipment. See §2.
•
Do not use mobile telephones or other radio, shortwave,
or microwave equipment in the vicinity of the equipment.
This kind of equipment can cause disturbances.
•
Only use the accompanying accessories that are supplied
by GymnaUniphy. See §7.6.
Other accessories can lead to an increased emission or a
reduced immunity.
•
Move the US head evenly over the skin during the
treatment. This prevents internal burns.
•
The US treatment heads are exchangeable. The device
detects the characteristics and supplies the right power
at the right frequency.
•
Handle the US heads carefully. With rough handling, the
characteristics can change. Test the US head if it falls on
the ground or knocks against something. See §5.1.1.
•
Check the US head at least once a month. During the
check, look for dents, cracks and other damage that could
allow liquids to ingress. Check whether the insulation of
the cable is still intact. Check whether all pins are present
and straight in the connectors. Replace the US head if the
head, the cable or the connector is damaged. See §5.1.
Summary of Contents for PULSON 200
Page 1: ...PULSON 200 User Manual...
Page 2: ......
Page 9: ...Pulson 200 10...
Page 31: ...Pulson 200 32...
Page 35: ...Pulson 200 36...
Page 39: ...Pulson 200 40...
Page 47: ...Pulson 200 48...
Page 49: ...Pulson 200 50...
Page 54: ...Pulson 200 55...
Page 55: ...Pulson 200 56...