PSM-SERIES PROGRAMMABLE POWER SUPPLY 

USER MANUAL 

 

       

 

 

18

z

 

I SET : Output current setting 

The minimum resolution of output current is at 1mA(output

10A at 

10mA). When set output to ON, press [I SET] to proceed current setting, 

the meter will display measurement value (the symbol of “

” is on with 

blinking). Besides, when the operation is set to “DISPLAY LIMIT” 

mode, the meter will display setting value (now the symbol of “

” is on 

without blinking). When the output is set to OFF, the output current 

setting can be done by pressing [I SET] directly or by changing the 

“DISPLAY LIMIT” mode (please refer to the operation of DISPLAY 

LIMIT). 

Setting method 1: Press [I SET][Number key (current value)] [ENTER] 

to set output current. 

Setting method 2: Press [I SET][number knob (voltage value)] to change 

the output current setting immediately, then press again [I SET] to 
complete current setting.

When use this method with the output on, the 

input current will be changed following the input value by rotating the 
knob which can be adjusted by moving the [

] or [

] cursor. 

For example: Set output voltage to 20.00A 

Press [I SET][2][0][.][0][0][ENTER] or [I SET][2][0][ENTER] 

PSM-SERIES PROGRAMMABLE POWER SUPPLY 

USER MANUAL 

 

       

 

 

19

When the current load runs through output terminal, if the current 

exceeds the setting value, the operation of the instrument will be set to 

Constant Current Mode (C. C. Mode), if the current does not exceed the 

setting value, the operation will be set to Constant Voltage mode (C. V. 

Mode). 

z

 

OVP SET: Over Voltage protection setting (LEVEL, ON/OFF 

/CLEAR) 

Setting method: After pressing [OVP SET], get into Over Voltage 

Protection Level setting, now use the number key (or knob) to input 

setting value, then press [ENTER] to complete the setting change. 

When get into Over Voltage protection state setting by selecting Over 

Voltage protection state and pressing [ENTER], use the knob to select 

ON/OFF/CLEAR, then press [ENTER] to complete the setting change. 

Remark: When the Over Voltage protection of the instrument is 

triggered, use the function of Over Voltage Protection 

Clear (OVP CLEAR) to clear its protection state. 

For example: Set Over Voltage protection value to 21.00V at ON state. 

After pressing [OVP SET] to get into OVP LEVEL setting, then input 

[2][1][.][0][0][ENTER] to get into OVP STATE setting, now set the 

knob select state to ON and press [ENTER] to complete setting. 

Summary of Contents for PSM-2010

Page 1: ...SUPPLY USER MANUAL i CONTENTS PAGE 1 PRODUCT INTRODUCTION Programmable Power Supply PSM 2010 3004 6003 USER MANUAL GW INSTEK PART NO 82SM 60030MA PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL ii IS...

Page 2: ...cification equipment and maintenance procedures at any time without notice Good Will Instrument Co Ltd No 7 1 Jhongsing Rd Tucheng City Taipei County 236 Taiwan 1 1 1 DESCRIPTION 1 1 2 FEATURE 2 2 TEC...

Page 3: ...LE POWER SUPPLY USER MANUAL vi SAFETY TERMS AND SYMBOLS These terms may appear in this manual or on the product WARNING Warning statements identify condition or practices that could result in injury o...

Page 4: ...oured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red If in doubt consult the instructions provided with the equ...

Page 5: ...1 2001 Voltage Dips Interrupts EN 61000 4 11 1994 A1 2001 Safety Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC IEC EN 61010 1 2001 PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 1 1...

Page 6: ...the 1 2 rack size reduction volume design 15 Save and Recall function of 100 groups setting and proceed together with auto operation can achieve the purpose of auto test 16 IEEE 488 2 and SCPI compati...

Page 7: ...MABLE POWER SUPPLY USER MANUAL 5 Specification PSM 2010 PSM 3004 PSM 6003 Voltage 0 02 1mV Current 0 1 1mA Memory Store Recall points 0 99 Temperature Coefficient per of output offset Maximum change i...

Page 8: ...To avoid electrical shock the power cord protective grounding conductor must be connected to ground When line voltages are changed replace the required fuses shown as below Model Line voltage Range Fu...

Page 9: ...M SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 9 15 DELAY Set the voltage and current output time in the automatic operation mode 16 AUTO RANGE AUTO INFO Proceed setting step by step During setting pr...

Page 10: ...SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 11 4 3 Display brief 1 Adrs Set the power supply to the address of Listen or Talk by using the interface card 2 Rmt Set the power supply to Remote Control...

Page 11: ...on mode when the meter displays the measurement value the symbol of becomes blinking 10 CC The CC light is on to indicate the instrument is in the CC mode 11 CV The CV light is on to indicate the inst...

Page 12: ...unless the LOCAL key is pressed at this time the OUTPUT is still working Whenever the power is reset the output will be at OFF status and the operation is through front panel operation mode z The ser...

Page 13: ...e function keys to proceed function setting after setting press again the function key to leave the function or press another key for another setting z V SET Output voltage setting The minimum resolut...

Page 14: ...o 20 00A Press I SET 2 0 0 0 ENTER or I SET 2 0 ENTER PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 19 When the current load runs through output terminal if the current exceeds the setting value th...

Page 15: ...LEAR to clear its protection state For example Set over current protection value to 21 00A at ON state After pressing OCP SET get into OCP LEVEL setting input 2 1 0 0 ENTER now set the knob select sta...

Page 16: ...Delay time to 9 minutes and 9 9 seconds Press SHIFT DELAY to appear MIN setting item input 9 and ENTER then select SEC setting item and input 9 9 ENTER to complete DELAY setting PSM SERIES PROGRAMMABL...

Page 17: ...ave the setting data to one of the memory groups STORE setting method Press the function of STORE use knob or number key to input the memory address to save the data For example Store the setting date...

Page 18: ...IB communication interface and set the ADDRESS to 08 Press SHIFT GPIB RS 232 use wheel knob to select GPIB and press ENTER to confirm the setting Then use wheel knob to select 08 and press ENTER to co...

Page 19: ...key to lock up panel setting Press the LOCK key to lock up the panel setting now the Output key is still working Press the key again for several seconds to unlock the panel setting PSM SERIES PROGRAM...

Page 20: ...ely PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 31 5 3 Operation Method The applied Voltage Current Unit for this series instruments is Volt and Amp z Constant Voltage Operation 1 Connect load to...

Page 21: ...the resolution PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 33 4 Set the desired output current value Press I SET the panel operation is set to current value input mode change to the desired value...

Page 22: ...tion level Over Voltage protection status Over Current protection status and the Delay time PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 35 2 Store the present setting status to the memory bank Pr...

Page 23: ...ng and store it to the memory bank group by group up to 100 groups maximum 3 Recall range setting Auto running operation range Regarding the setting description please refer to AUTO RANGE operation 4...

Page 24: ...2 Sampling and wiring application for the output terminal of European Regulation 1 CV Regulation When the voltage drop between the output terminal and positive sensing point increases 1V its voltage...

Page 25: ...lue of the output setting Model PSM 2010 PSM 3004 PSM 6003 Output Range Item 8V 20A 20V 10A 15V 7A 30V 4A 30V 6A 60V 3 3A Output Voltage 8 240V 20 60V 15 450V 30 900V 30 900V 61 800V Output Current 20...

Page 26: ...less you are qualified to do so 6 1 Fuse Replacement If the fuse blows the display will not light and the power supply will not operate The fuse should not normally open unless a problem has developed...

Page 27: ...on transient response time and open test CV voltage calibration Measurement Techniques Assure a precise Load regulation Voltage peak to peak and transient response time the test joint point must be co...

Page 28: ...ibrated point input DMM measurement value then press ENTER Note The DMM selected for the measurement must have the resolution of 4 digits after the decimal point at least and take three digits after t...

Page 29: ...ing into Current calibration procedure PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 49 Step 5 1 Firstly proceed the calibration procedure of LO calibrated point input DMM measurement value then pr...

Page 30: ...al short Switch to SAVE by using the knob and press ENTER to leave the mode with storing Also can press SHIFT to leave the mode without storing PSM SERIES PROGRAMMABLE POWER SUPPLY USER MANUAL 51 Step...

Page 31: ...e ground Digital Control circuit MPU Micro Processor Unit VFD DISPLAY Interface Control Card GPIB IEEE 488 and RS 232 Power output circuit DAC Digital to Analog Converter ADC Analog to Digital Convert...

Reviews: