MULTI-OUTPUT POWER SUPPLY 

                                          USER MANUAL 

 

 

18 

(4) CH3 Power Supply Operation 

The CH3 supply provides 2.2~5.2V(GPS-4302/4303) and 3~6V(GPS-4251) 

DC output with 3 amps (GPS-3303&4302), 1 amp (GPS-4303) and 2.5 amp 

(GPS-4251) current capacity. The supply is ideal for us with TTL circuits 

(GPS-3303 5V Fixed). 

A.

 

Turn off the power supply and the equipment to be powered during 

hook-up. 

B.

 

Connect the positive polarity of the device being powered to the red (+) 

terminal of the CH3 supply. 

C. Connect the negative polarity of the device being powered to the black (-) 

terminal of the CH3 supply. 

D.

 

If the red OVERLOAD indicator lights, means exceeding load has been 

placed on the supply, it will cause voltage and current to drop and interfere 

the proper operation of the CH3 supply. To correct this situation, the load 

on the supply must be decreased so that no more than 3 amps 

(GPS-3303&4302), 1amp (GPS-4303) and 2.5 amps (GPS-4251) of 

current are required. 

 

(5) CH4 Power Supply Operation 

The CH4 supply provides an 8~15V(GPS-4302/4303/4251) DC output with 

a 1 amps current capacity. 

A.Turn off the power supply and the equipment to be powered during 

hook-up. 

B.

 

Connect the positive polarity of the device being powered to the red (+) 

terminal of the CH4 supply. 

C. Connect the negative polarity of the device being powered to the black (-) 

terminal of the CH4 supply. 

D.

 

If the red OVERLOAD indicator lights, exceeding load has been placed on 

the supply, it will cause voltage and current to drop and interfere proper 

operation of the CH4 supply. Under this situation, the load on the supply 

must be decreased so that no more than 1 amp of current is required. 

MULTI-OUTPUT POWER SUPPLY 

                                          USER MANUAL 

 

 

19

(6) Dynamic Load Operation & Application 

A. When select to dynamic load position, the max peak current is at 1.7 times 

rating current. The features are only applied for audio circuit of amplifier 

and audio production lines. Change the position of wafer J111 of CH1 and 

J309 of CH2 from “OFF” to “ON”. Please refer to Fig.6-1 Adjustment 

Location. 

B. For other application and testing (Safety or CE. etc.), must set the wafer at 

“OFF” position 

(7)

 

Output ON/OFF Action   

The output ON/OFF action is controlled with a single control, the output 

switch is pushed on, a high single output is on and output LED is on, while 

the output switch is pushed off, or tracking switch is pressed, output will 

disable.  

 

CAUTION: The output terminals are for use only with the equipment 

which has no accessible live parts.    The output terminals should not 

be connected to any hazardous live parts. 

 

(8)

 

Fan Control 

1)

 

The fan of the power supply will not work upon power on until the 
temperature of the heat sink rises up to 32

℃±

5

 after adding load 

to output terminal. The more the temperature of the heat sink rises, 

the more the rolling speed of the fan gets fast. The fastest rolling 
speed is when the temperature reaches to 70

2)

 

To avoid damaging the power supply, if the fan fails to work when 

the temperature reached to the appropriate value, turn off the 

instrument and check the cause. 

Summary of Contents for GPS-2303

Page 1: ...A 0 2 3mA 0 2 3mA 3mArms 0 5 10mV of the CH1 300mV 0 01 3mV 2mVrms 5Hz 1MHz 5mV 15mV 2mVrms 5Hz 1MHz GPS 4303 1A GPS 3303 3A GPS 4251 2 5A 5mV 10mV 2mVrms 5Hz 1MHz 1A 3 digits 0 5 LED display GPS 4303...

Page 2: ......

Page 3: ...rective 73 23 EEC amended by 93 68 EEC EN 61010 1 2001 IEC 61010 1 2001 MULTI OUTPUT POWER SUPPLY USER MANUAL ii Declaration of Conformity We GOOD WILL INSTRUMENT CO LTD No 7 1 Jhongsing Rd Tucheng Ci...

Page 4: ...c 5 4 Operation Mode 11 11 11 12 13 6 MAINTENANCE 6 1 Fuse Replacement 6 2 Line Voltage Conversion 6 3 Adjustments 6 4 Cleaning 19 19 19 20 24 iv SAFETY TERMS AND SYMBOLS Please take a moment to revie...

Page 5: ...package GPS 4302 GPS 4303 and GPS 4251 consist of four identical independent adjustable DC power supplies while two for GPS 2303 and three for GPS 3303 A front panel switch selects one of three opera...

Page 6: ...n 0 01 3mV Load regulation 0 01 3mV rating current 3A Load regulation 0 02 5mV rating current 3A MULTI OUTPUT POWER SUPPLY USER MANUAL 3 3 Recovery time 100 s 50 load change minimum load 0 5A 4 Ripple...

Page 7: ...m function description Single phase input power is applied to transformer through the input circuit Auxiliary rectifier D1021 1024 provides a bias voltage filtered by capacitor C103 C104 for the prere...

Page 8: ...MULTI OUTPUT POWER SUPPLY USER MANUAL 6 Figure 3 1 Block Diagram MULTI OUTPUT POWER SUPPLY USER MANUAL 7 4 PANEL CONTROLS AND INDICATORS Fig 4 1 Front Panel...

Page 9: ...king operation 7 Current Control Adjust the output current of the CH1 supply and as the adjustment control for the maximum output current of the CH2 supply when either parallel or series tracking oper...

Page 10: ...GPS 2303 3303 28 Output switches ON OFF the output MULTI OUTPUT POWER SUPPLY USER MANUAL 11 29 30 TRACKING Mode Switch Two push button switches that select Independent mode series tracking mode or par...

Page 11: ...oltage Constant Current Characteristics The working characteristic of these series Power Supplies is called a constant voltage constant current automatic crossover type This permits continuous transit...

Page 12: ...CH2 supply voltage is automatically set to the same as the CH1 supply by using the CH1 VOLTAGE controls A Set the power supplies to the Tracking series mode by engaging the left Tracking switch and re...

Page 13: ...allows for a rating voltage supply with a double rating current capability Only the CH1 output terminals are used for parallel tracking operation In the parallel tracking mode the CH2 supply output vo...

Page 14: ...he supply it will cause voltage and current to drop and interfere proper operation of the CH4 supply Under this situation the load on the supply must be decreased so that no more than 1 amp of current...

Page 15: ...been made in a circuit affecting adjustment accuracy or if you have a reason to believe the unit is out of adjustment The recommended calibration device is a multimeter with an accuracy of 0 1 DCV or...

Page 16: ...circuit board VOLTAGE CURRENT CONTROL potentiometer board until the voltage read from the multimeter is the same as it was MULTI OUTPUT POWER SUPPLY USER MANUAL 23 across the CH1 output terminals Retu...

Page 17: ...ff by adjusting the trimmer potentiometer VR803 from the CH4 voltage indicator circuit board located on the meter or A D converter board for a reading of rate voltage on the CH4 C Turn VR601 on the ma...

Page 18: ...MULTI OUTPUT POWER SUPPLY USER MANUAL 26 Fig 6 1 Adjustment Location...

Reviews: