Adjustment Procedure (Manual procedure) 

Model no. 

GDM

–825XA 

Document No.  

TO 

– XX -XXX 

Page 1 of  

NO. 

Item 

Condition 

 

66 

ACV 

750V 

750V 

1kHz 

Use the 2A input and the COM terminal for DCmA or ACmA calibration.

 

41 

DCmA 

0A 

0A 

 

 

42 

DCmA 

10mA 

10mA 

 

 

43 

DCmA 

100mA 

100mA 

 

 

21 

ACmA 

1.9mA 

1.9mA  1kHz 

22 

ACmA 

19mA 

10mA  1kHz 

23 

ACmA 

100mA 

100mA  1kHz 

Use the 10A input and the COM terminal for DCA or ACA calibration.

 

31 

DCA 

0A 

0A 

 

 

32 

DCA 

10A 

10A 

 

 

51 

ACA 

1.9A 

1A 

1kHz 

52 

ACA 

10A 

10A 

1kHz 

Appendix 2 

4WR 

Connection 

a. 4WR is a connection method for 4 wires. 

b. When 4WR is performed, EX SNS in Fluke5720A should be turned 

ON

 

c. The 4WR connection method is shown as the diagram below. 

 