Adjustment Procedure (Manual procedure)
Model no.
GDM
–825XA
Document No.
TO
– XX -XXX
Page 1 of
NO.
Item
Condition
66
ACV
750V
750V
1kHz
Use the 2A input and the COM terminal for DCmA or ACmA calibration.
41
DCmA
0A
0A
42
DCmA
10mA
10mA
43
DCmA
100mA
100mA
21
ACmA
1.9mA
1.9mA 1kHz
22
ACmA
19mA
10mA 1kHz
23
ACmA
100mA
100mA 1kHz
Use the 10A input and the COM terminal for DCA or ACA calibration.
31
DCA
0A
0A
32
DCA
10A
10A
51
ACA
1.9A
1A
1kHz
52
ACA
10A
10A
1kHz
Appendix 2
4WR
Connection
a. 4WR is a connection method for 4 wires.
b. When 4WR is performed, EX SNS in Fluke5720A should be turned
ON
c. The 4WR connection method is shown as the diagram below.