GDM-8246 MULTIMETER
PROGRAMMER MANUAL
⎯
22
⎯
Examples:
*ESE 65 sets the ESER to binary 0100 0001.
If the ESER contains the binary value 1000 0010, the *ESE? will return
the value of 130.
*ESR?
(query only)
Function:
Return and clear the contents of the Standard Event Status Register
(SESR).
Syntax:
*ESR?
Returns:
<NR1> is a number from 0 to 255 that indicates the decimal value of the
binary bits of the ESER.
Examples:
If the ESER contains the binary value 1100 0110, the *ESR? will return
the value of 198.
*IDN?
(query only)
Function:
Return the unique identification code of the DMM.
Syntax:
*IDN?
Returns:
<string> includes Manufacturer, model number, serial number and
firmware version.
GDM-8246 MULTIMETER
PROGRAMMER MANUAL
⎯
23
⎯
Examples:
*IDN? Returns
GW_Inc, GDM-8246, FW1.00
*OPC
Function:
The command form (*OPC) sets the operation complete bit (bit 0) in the
Standard Event Status Register (SESR) when all pending operations are
finished.
The query form (*OPC?) tells the Digital Multimeter to place an ASCII 1
in the Output Queue when the DMM completes all pending operations.
Syntax:
*OPC
*OPC?
Returns:
1
*RST
(no query form)
Function:
Set all control settings of DMM to 1000V DCV range but does not purge
stored setting. The equivalent panel control will be set as below:
*SRE
Function:
Set the contents of the Service Request Enable Register (SRER). The
query form returns the contents of the SRER. Bit 6 of the SRER is
always zero. The bits on the SRER correspond to the bits on the SBR.
Syntax:
*SRE <NR1>