Trouble Shooting
The Model 12V500 provides internal diagnostics that monitor proper simula-
tor operation. If a monitored system detects a condition that is outside of the
normal operation parameters, the simulator will deactivate the solution and head
space heaters as well as the agitator motor. The simulator will emit an audible
alarm and display the detected condition code number on the LCD followed by a
message and icon detailing the condition. The LCD will alternate displaying the
status code and message until the condition is corrected or the simulator is reset.
The audible alarm will stop after 1 minute.
Resetting The Simulator
Turning the power off and on will reset the simulator.
Status Codes and Messages
The following list of status codes and messages indicate the cause and corrective
action:
11
PROBLEM:
No solution in container or simulator power has been turned on
with top housing not attached to the container.
REMEDY:
Turn the simulator off. Fill container with 500 ml of certified solu
-
tion and reassemble top housing and container.
PROBLEM:
The temperature sensor for the heater control circuit is either
open or shorted.
REMEDY:
Cannot be remedied in the field. Requires service.*