Method of cleaning and disinfection
Cleaning
We recommend that any parts of the mobile lifter that users/helper may come
in contact with are cleaned with Novadan Des Wipes, Novadan Disinfect O or
similar disinfection product.
Novadan Des Wipes:
Use a disposable wipe for cleaning and disinfection of all types of surfaces,
machines and instruments in the healthcare sector.
NB:
Do never clean the lifter with strong acid, base or alcohol. Never clean
the lifter in an autoclave machine.
7 .00
Troubleshooting
Error:
Lifter does not react to input from the buttons on the hand control.
1 . Does the green or yellow LED illuminate on the lifter when the control
buttons on the hand control are activated?
Yes see point 6.
No no light visible – see point 2.
No only red light visible – see point 3.
No yellow service light visible or audible “Beep”– see “indicator and audio
signals matrix”
2 . Is the battery installed in the lifter?
Yes see point 4.
No install a fully charged battery.
3. Is a battery with sufficient charge installed in the lifter?
No install a fully charged battery.
4 . Is the emergency stop activated?
Yes turn emergency stop in the direction of the arrow to allow it to release
and press any button on the hand control.
No see point 5
5. Is the connector for the hand control fitted in the lifter, and is the lifter
reacting to the buttons located on the control box?
Yes replace hand control
No fit connector