17
©
Guldmann G
B
/U
S
-1
0/
20
19
•
#
558350_1
©
Guldmann G
B
/U
S
-1
0/
20
19
•
#
558350_1
6 . Are the connectors for the lifting motor and/or leg adjustment motor
fitted?
Yes see point 5
No fit connector
Contact the Guldmann Service Team if the fault cannot be found and
corrected.
8 .00
Service and lifespan
According to the international standard EN/ISO 10535 “Lifter for the transfer
of disabled persons - Requirements and test methods” a safety inspection of
the lifter must be performed at least once a year.
The lifter has an expected life span of 10
years. The lifespan is estimated on the basis
of correct use, cleaning and maintenance, plus
annual service and maintenance carried out by
qualified service engineers of the Guldmann
Service Team. At the end of the expected lifes-
pan, the lifter must then be assessed by quali-
fied service staff as to the lifters future use. The
lifespan of the lifting actuator is determined by
how regular the lifter is used.
No part of the equipment shall be serviced when in use with a patient.
The lifter is controlled by a microprocessor PC board which can be damaged
if it is being touched without the necessary precautions. Therefore the electro-
nics must only be serviced by qualified Guldmann Team.