25
Operation
Operating mode
Warning
1. Pre-Disinfection or
Decontamination
- Soak immediately just after
usage all instruments in a
disinfectant solution combined
with proteolytic enzyme if
possible.
- Follow instructions and observe
concentrations and immersion time given by
the manufacturer (an excessive concentration
may cause corrosion or others defects on
instruments). - The disinfectant solution
should be aldehyde free (to avoid blood
impurities fixation). - Do not use disinfectant
solution containing Phenol or any products
which are not compatible with the instruments
(See general recommendation). - For visible
impurities that are observed on instruments,
a pre-cleaning is recommended by brushing
them manually with soft material.
2. Rinsing
- Rinse manually and abundantly
the accessories with current
water
3. Manual Cleaning
- clean manually the accessories
with an adequate brush,
preliminary soaked in a clean
pre-disinfectant solution
- the file clip mechanism has to be activated
during the cleaning process (press several
times the push button) - no visible impurities
should be observed on the accessories
4. Rinsing
See point # 2
5. Disinfection
- Immerse the accessories
in a disinfectant solution
(bactericidal, virucidal,
fungicidal, tuberculocidal and
aldehyde free) according to the
manufacturer recommendations
- Follow instructions and observe
concentrations and time given by the
manufacturer
6. Final rinsing
- See point # 2 - After rinsing,
the accessories have to be dried.