11
Picture 9
b) Insert the tester.
c) After powered on, If the indicating bar indicates within ±3 bars away from the
dial 0.5, the device functions normally [Picture 9]. If the indicating bar is outside
the range, the device cannot measure accurately. On this occasion, please contact
authorized distributor or manufacturer for help.
3.3 Connect to compatible Endo Motor. (Optional)
3.3.1Plug one end of the USB line into the USB socket on the right side of
the device, and connect the other end with compatible Endo Motor as shown in
Picture 10(a). There is no difference between those two ends. As shown in Picture
10 (b), when the Contra-angle icon is lit, the Apex locator and Endo Motor can
communicate normally, so that the 2-in-1 function can be realized in Endo Motor.