2.10 Please recharge the battery when low battery indicator flashes.
2.11 Please use original components, the components made by other companies
may cause inaccurate measurement or un-measurable.
2.12 Avoid the connection between the outside and inside liquid of endodontic
during measuring in order to avoid the measuring difference.
2.13 Keep endo file and file clip away from any other metal or instruments.
2.14 To ensure that short circuits do not impair the measurements, be particularly
careful with patients fitted with metal crowns or bridges. Please confirm the wetness
of the endo to ensure the reliability of the measuring. If it is confirmed that the endo
file hasn't reached the apex yet the data showed on the apex locator is too low, please
check whether the endo is too dry and confirm it with X-ray.
2.15 This device have electromagnetic interference, the patient or doctor who
with a heart pace maker are forbidden to use this device and the device is susceptible
to other device which produces electromagnetic interference. Dentists should be
cautious about operation under such environment.
2.16 The guarantee is valid for normal usage conditions. Any disassembly will
render the guarantee void, the professionals of Woodpecker company will offer the
repair service during guarantee period.
2.17 Any modification will render the guarantee void and may cause harm to the
patient.
2.18 Only the original adapter and lithium battery could be used to this machine.
2.19 Not to position equipment to make it diffcult to operate the disconnection
device.
2.20 The adapter must be connected to an appropriate power source in the
instructions.