27
Automated Cleaning:
Use a washer-disinfector meeting the requirements of the ISO 15883 series. Put
the instrument into the machine on a tray. Connect the instrument with the WD by
using suitable adapter and start the program:
• 4 min pre-washing with cold water (<40°C);
• emptying
• 5 min washing with a mild alkaline cleaner at 55°C
• emptying
• 3 min neutralising with warm water (>40°C);
• emptying
• 3 min neutralising with warm water (>40°C);
• emptying
• 5 min intermediate rinsing with warm water (>40°C)
• Emptying
The automated cleaning processes have been validated by using 0.5% neodisher
MediClean forte (Dr. Weigert). Note Acc. to EN ISO 17664 no manual reprocessing
methods are required for these devices. If a manual reprocessing method has to be
used, please validate it prior to use.
6.6 Disinfection:
Automated Thermal Disinfection in washer/disinfector under consideration of
national requirements in regards to A0 value (see EN 15883). A disinfection cycle of
5 min disinfection a 93°C has been validated for the device to achieve an A0 value
of 3000.
6.7 Automated Drying:
Drying of outside of instrument through drying cycle of washer/ disinfector. If
needed, additional manual drying can be performed through lint free towel. Insufflate