7. Error States of the device

Device  continuously  checks  its  state  during  operation.  In  case  error  is  found  LCD  displays

corresponding error code: 

Error 0 

- first  line  of  LCD  displays  „Err0“.  Check  sum  error  of  stored  setting  inside  device’s 

memory. This error appears if incorrect writing  procedure to device’s memory occurred or if 
damage of calibration data appeared. At this state device does not measure and calculate values. 
It is a serious error, contact distributor of the device to fix. 

Error 1 

- measured or calculated value except in pressure is over upper limit of allowed full scale 

range. There is a reading „Err1“ on LCD display. Value read from the device is +999.9.  This 
state appears in case of: 

Measured  temperature  is  higher  than  approximately  600°C  (i.e.  high  non-measurable
resistance of temperature sensor, probably opened circuit).

Relative humidity is higher than 100%, i.e. damaged humidity sensor, or humidity calculation
of humidity is not possible (due to error during temperature measurement).

Computed  value  –  calculation  of  the  value  is  not  possible  (error  during  measurement  of
temperature or relative humidity or value is over range).

The value of pr999.9 hPa is correct value! 

Error 2 - t

here is a reading „Err2“ on LCD display. Measured or calculated value is below lower 

limit of allowed full scale range. Value read from the device is -999.9. This state appears in case 
of: 

Measured temperature is lower than approximately -210°C (i.e. low resistance of temperature
sensor, probably short circuit).

Relative  humidity  is  lower  than  0%,  i.e.  damaged  sensor  for  measurement  of  relative
humidity,  or  calculation  of  humidity  is  not  possible  (due  to  error  during  temperature
measurement).

Measured pressure value 

with added correction for altitude

 is outside of range from 300 hPa

to 1350 hPa (from 4.351 PSI to 19.580 PSI) or the pressure sensor is damaged. Please check
setting of pressure correction for altitude with User’s software.

Computed value – calculation of computed value is not possible (error during measurement
of temperature or relative humidity).

Error 3 - 

There  is  a  reading  „Err3“  on  LCD  display  upper  line.  Error  of  internal  A/D  converter 

appeared (converter does not respond, probably damage of A/D converter). This error does NOT 
affect  pressure  measurement.  Rest  values  are  NOT  measured.  It  is  a  serious  error,  contact 
distributor of the device. 

Error 4 – 

there is a reading „Err4“ on LCD display. It is internal device error during pressure sensor 

initialization. Under this condition device does NOT measure atmospheric pressure. Value read 
from  device  is  -999.9.  Pressure  sensor  is  probably  damaged.  It  is  a  serious  error,  contact 
distributor of the device. 

Error 5,  6 - 

there  is  problem  with  assigned  value  to  output  relay,  there  is  some  wrong  setting 

(mismatch). This error appears if incorrect writing procedure to device’s memory occurred. 

Error 9 - 

inserted  password  is  not  valid,  there  is  shown  for  short  time  message  “Err9”  on  LCD 

display. 

Summary of Contents for AR006687

Page 1: ...Temperature humidity and pressure regulator with RS485 serial output AR006687 AR006689 AR006690 AR006688 AR006693 Instruction Manual...

Page 2: ...lator 19 6 1 4 Far controlling of output relays status 20 6 1 5 Output relay alarm condition configuration with using of Modbus communication protocol 2 21 6 2 PROTOCOL COMPATIBLE WITH ADVANTECH ADAM...

Page 3: ...ut the cover Do not use the instrument if it does not work correctly If the instrument seems not to work correctly have it checked by a qualified service person Do not use the instrument in an explosi...

Page 4: ...o Modbus RTU communication protocol If you would like to use different communication protocol then Modbus RTU it is necessary preset them see latter Temperature relative humidity and pressure sensors...

Page 5: ...as far as possible from potential interference sources For devices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor...

Page 6: ...6 3 Typical application wiring connection of terminals 4 Dimensions AR006687...

Page 7: ...7 AR006689 AR006693...

Page 8: ...8 AR006688 AR006690 Probe for compressed air measurement...

Page 9: ...tact and hysteresis for return to open status For binary inputs if it is preset to HI then output relay is closed when input contact is opened if it is preset to Lo then output relay is closed when in...

Page 10: ...information about Relay1 and then Relay2 setting step by step Information about Relay1 are indicated with symbol for Relay2 with symbol at right top corner of LCD It is possible to edit parameters af...

Page 11: ...n measured value decrease under high limit value minus the hysteresis value See example High limit HI Delay s Hysteresis Low limit Lo the relay closes when measured value falls below preset low limit...

Page 12: ...h the software 5 5 2 Setting with using regulator keys If Actual regulator preset is displayed see chapter Actual regulator preset at page 10 it is possible to enter into edit mode and edit displayed...

Page 13: ...ount four screws in the corners of the box and remove the lid 3 closed jumper placed nearby keys 4 press and keys together and keep them down 5 connect power supply to regulator you should see message...

Page 14: ...elay2 stays unchanged in previous status Relay1 switch OFF Relay1 switch On stays unchanged in previous status Relay2 switch OFF Relay2 switch On Change of Password for setting through keys PASS This...

Page 15: ...ation of inputs is done by three LED diodes This option defines at which binary input status should LED diode indication lights LED diode lights if binary input is OPEN or CLOSED LED lights if binary...

Page 16: ...le to set corresponding pressure offset for this altitude It is possible to change it after SET key pressing Pressure correction value is then automatically added to measured pressure value See chapte...

Page 17: ...line see Computed value selection below It is possible to change it after SET key pressing show value on LCD don t show value on LCD Show barometric pressure at LCD display This item provides to choic...

Page 18: ...re pressure by device type value shown at second LCD line relative humidity dew point temperature temperature unit C pressure unit hPa correction for altitude 0 hPa absolute pressure computed value pr...

Page 19: ...on of jumper is as follows Jumper opened device memory is protected from writing from device side it is only enabled to read measured value writing to memory is disabled no change of settings is enabl...

Page 20: ...address 0x31 or 0x30 Status word 16b value return bite description Bit0 0 1 jumper open closed Bit1 unused Bit2 0 always 0 Bit3 0 1 Relay1 open closed Bit4 0 1 Relay2 open closed Bit5 0 1 internal aco...

Page 21: ...d into device memory confirm change Then registers 0x004F 79 and 0x0044 68 are automatically cleared zero It automatically unlocks the keyboard too Whole setting process can be cancelled by writing va...

Page 22: ...on examples are listed in a separate document entitled Description of communication protocols 6 2 Protocol compatible with Advantech ADAM standard Control units communicate on master slave principle i...

Page 23: ...means manufacturer setting of communication protocol is done For communication with device which measure more than one value there is necessary to add at the end of command number of communication cha...

Page 24: ...appears in case of Measured temperature is lower than approximately 210 C i e low resistance of temperature sensor probably short circuit Relative humidity is lower than 0 i e damaged sensor for meas...

Page 25: ...nly value number without corresponding unit 9 Preventive maintenance and calibration Pay heed to recommended calibration interval Do not subject the device to mechanical stress 10 Technical specificat...

Page 26: ...emperature compensated at entire temperature range Accuracy 2 5 RH from 5 to 95 RH at 23 C 73 4 F Range 0 to 100 RH Resolution 0 1 RH Measuring temperature and humidity range is limited in accordance...

Page 27: ...emperature probe model Resolution 0 1 C 0 2 F Accuracy 0 2 C 0 4 F without probe 10 1 7 AR006693 temperature relative humidity and barometric pressure regulator Temperature and relative humidity param...

Page 28: ...temperature Range 60 to 80 C 76 to 176 F Accuracy 1 5 C 2 7 F at ambient temperature T 25 C 77 F and relative humidity RH 30 for more details see graphs below Absolute humidity Range 0 to 400 g m3 Acc...

Page 29: ...dity and temperature restriction 10 2 Operating conditions Operating temperature and relative humidity electronics 30 to 80 C 22 to 176 F 0 to 100 RH without condensation probe AR006690 30 to 105 C 22...

Page 30: ...over downwards Not allowed manipulations It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically ag...

Page 31: ...nd pin 8 B The pull up pull down and termination resistors are part of the transmitter These internal resistors can be connected to the bus by connecting the corresponding pins of CANON connector for...

Page 32: ...g valve 3 probe 4 flow chamber SH PP 5 closing valve 6 outlet tube sampling 1 end of the tube placed in the centre of pressure pipelines distribution of moisture in the pipe cross section is not homog...

Page 33: ...ins basic information about connection of the detectors optional accessories to binary inputs of regulator The detectors install in accordance with the attached instructions for use SP008 power detect...

Page 34: ...is used connect the terminals ALARM to the terminals GND and IN1 If the detector is powered from the regulator it is necessary to use power supply of 12Vdc for power supplying the regulator Notice Th...

Reviews: