22
GMG-200
310-2310-001 EN20
ERRORS
11 Errors
TIPS AND RECOMMENDATIONS
When the strap is completely retracted, the display may indicate a force > 0
since the strap limit switch presses against a buffer spring!
Possible error indications
►
Remedy common errors by following the instructions in the display!
Display
Description
Remedy
High force
Start: pull in
Permitted motor current
exceeded
Start strap retraction
Winding - error
Start: pull in
Internal traction ca-
ble winding error
Extract the strap and then take care that
the strap is drawn in straight and evenly
when attempting again
EEPROM - error
Start: pull in
Inconsistency in RAM and/or
EEPROM data
Start strap retraction
Counter reading error:
Measurement results are not
displayed
Measurement invalid - repeat measur-
ing run
Blank display - device is
in self-discharge protection
mode:
The device has an automatic
shutdown feature to prevent
deep discharge. If within ten
minutes no functions occur
and no data transmitted, the
device switches itself off
Press “Start” to switch on the device