GSI AudioStar Pro™ Clinical Audiometer
36
D-0100778 Rev C
The selection of any stimulus will deselect a previously selected stimulus on the
opposite channel if the stimuli are not compatible. Refer to the following table
for the stimuli compatibilities listing:
Valid Stimuli Combinations
Channel 1 Stimulus
Tone
Mic
Ext. A
Ext. B
NB
Noise
S Noise
White
Noise
Channel 2
Stimulus
Tone
Valid
Valid
Valid
Valid
Valid
Invalid
Valid
Mic
Valid
Valid
Valid
Valid
Invalid
Valid
Valid
Ext. A
Valid
Valid
Valid
Valid
Invalid
Valid
Valid
Ext. B
Valid
Valid
Valid
Valid
Invalid
Valid
Valid
NBNoise
Valid
Invalid
Invalid
Invalid
Valid
Invalid
Invalid
S Noise
Invalid
Valid
Valid
Valid
Invalid
Valid
Invalid
W Noise*
Valid
Valid
Valid
Valid
Invalid
Invalid
Valid
NOTE:
If White Noise is selected on both channels, then calibration is set to
speech effective masking levels. If White Noise is selected on one channel only,
calibration will be set to mask the stimulus type on the opposite channel.
Transducer Output Selector
The Transducer pushbuttons allow the easy selection of the transducer for each
stimulus available for Channel 1 and Channel 2. A transducer selection may be
changed at anytime.
Valid Transducer Combinations
Channel 1
Phone
Bone
Speaker
Insert
High
Freq.
Phones
Channel 2
Phone
Valid
Valid
Valid
Invalid
Invalid
Bone
Valid
Valid
Valid
Valid
Valid
Speaker
Valid
Valid
Valid
Valid
Valid
Insert
Invalid
Valid
Valid
Valid
Invalid
High Freq.
Phones
Invalid
Valid
Valid
Invalid
Valid