D-0100778 Rev C
11
Customer Responsibility
Warning!
This product and its components will perform reliably only when operated and
maintained in accordance with the instructions contained in this manual,
accompanying labels, and/or inserts. A defective product should not be used.
Make sure all connections to external accessories are snug and secured properly.
Parts which may be broken or missing or are visibly worn, distorted or
contaminated should be replaced immediately with clean, genuine replacement
parts manufactured by or available from GSI.
This product should not be used in the presence of fluid that can come into
contact with any of the electronic components or wiring. Should the user suspect
fluids have contacted the system components or accessories, the unit should not
be used until deemed safe by a GSI certified service technician.
Do NOT use in the presence of flammable gaseous mixtures. Users should
consider the possibility of explosions or fire when using this device in close
proximity to flammable anesthetic gases.
Do NOT use the AudioStar Pro in a highly oxygen-enriched environment, such
as a hyperbaric chamber, oxygen tent, etc.
Periodically, have a service technician perform electrical safety checks on the
unit in order to maintain continued compliance to IEC and UL 60601-1.
Equipment is not user repairable. Repairs and battery replacement must be
performed by a qualified service representative only. GSI will make available
any instructions and diagrams to repair devices that it deems appropriate to be
repaired in the field.
Elimination of Ambient Noise
The GSI AudioStar Pro may be installed in a single room environment or as part
of a two room
suite.
Excessive noise in the test environment, such as that produced by conversation,
office equipment, or printers, reduces test validity because it tends to mask the
test signals. This is especially true at the lower frequencies where earphone
cushions provide less effective attenuation. A room that attenuates sound may be
required if ambient noise at the patient’s ears reaches levels sufficient to cause
apparent hearing loss at the lower frequencies.
The following table shows the maximum background levels that can be present
inside the room while a valid hearing test is being conducted. These values apply
for hearing threshold measurements to 0 dB HL.