English (GB)
9
Fig. 4
Measuring cell AQC-D4
8. Installation
8.1 Transport and storage
8.2 Unpacking
• Check the scope of delivery.
• Install as soon as possible after unpacking.
8.3 Installation requirements
• The location must be vibration-free, dry, dust-free and free of
corrosive, pungent fumes or aggressive solvents.
• For safety reasons, the customer should install an earth
leakage circuit breaker for the measuring amplifier
Conex
®
DIA and the motor of the measuring cell.
• Maximum permissible cable length between measuring cell
and measuring amplifier: 3 metres.
• Ensure a continuous supply of sample water.
• Retain the permissible inlet pressure and counter-pressure of
the sample water.
– Fit a pressure booster pump or pressure reducer, if
necessary.
8.4 Mounting
The flow armature is fastened to the mounting plate from factory.
1. Drill four dowel holes with a diameter of 10 mm and minimum
60 mm deep. See section
6.2 Dimensional sketch / drilling
2. Mount the mounting plate on the wall using the mounting
material provided. Distance between the mounting plate and
the wall: minimum 20 mm.
Fig. 5
Wall-mounting guide for AQC-D4 preassembled
system
TM
03
71
25
45
06
Pos. Description
A
Location of pH and redox electrode
B
Location of sample water regulating device
C
Location of measuring and reference electrode
D
Location of temperature sensor Pt100
E
Measuring and reference electrode
F
Flow fitting
G
Limitation of sample water with ascending pipe
H
Cleaning drive
I
Sample water deficiency sensor
J
Connection for sample water inlet
K
Connection for sample water outlet
L
Sample water adjusting device
M
Counter-electrode
N
Contacting of counter-electrode
O
Screw
part with magnetic stirring stone
A
B C
D
H
O
M
K
G
F
E
L
I
J
N
Caution
Transport the device carefully, do not drop!
Store in dry conditions b5 °C and +40 °C.
Store electrodes between -5 °C and +30 °C.
Keep protective caps moist with 3-molar potassium
chloride solution.
Caution
Do not allow any foreign bodies to enter!
Caution
If you do not observe the installation requirements,
the device may be damaged!
The measurements may not be correct!
Caution
The measuring cell breaks when the screws are
tightened! Do not tighten the screws; only screw
them in without tightening them!
Caution
The potentiostatic measuring cell or preassembled
system must be mounted vertically!
Caution
Do not pinch the cable! Always observe the
wall-mounting guide below.
T
M
03
71
26
45
06
20 mm