Page 34
SFM 6.1 Slurry Flow Meter
DATA LOGGING (Optional)
Setup
Select
Data Logging
from Menu Selections.
Log Site ID
Enter a number from
00
to
99
. The site ID will become part
of the downloaded file name to help distinguish downloads
from different instruments. Press
to store the setting.
Mode
Select Velocity (e.g. ft/sec or m/sec).
Flow
(e.g. USGPM or
l/sec). Press
to store the setting.
Set Date
Press
or
to scroll and select Month, Day and Year.
Press
to store the setting.
Set Time
Press
or
to select the current time in Hours, Minutes and
Seconds. Press
to store the setting.
Interval
Press
or
to select the logging interval.
Press
to store the setting.
Log
Stop
,
Start
or
Delete
the log file.
Press
or
to
Delete
and
to delete the log file.
Press
or
to
Start
and
to start the logger.
Note:
You MUST delete old log and start a new log AFTER having set
changes to
Log Site ID
,
Mode
and/or
Interval
for those changes to be
applied to the log file.
RETRIEVE LOG FILE
Plug a USB Flash Memory Drive (not supplied by Greyline) into the USB output
cable from the instrument. The instrument display will show the USB file
download icon until the log file is transferred to the memory card and then display
file download completed icon. The USB flash drive may be removed.
Download file names will appear in this format:
Tag is set according to the
Log Site ID
entered in the instrument
Data Logging
menu.
Summary of Contents for SFM 6.1
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 SFM 6 1 Slurry Flow Meter CONNECTIONS ...
Page 7: ...Page 7 SFM 6 1 Slurry Flow Meter CALIBRATION MENU ...
Page 8: ...Page 8 SFM 6 1 Slurry Flow Meter ICONS ...
Page 19: ...Page 19 SFM 6 1 Slurry Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS ...
Page 28: ...Page 28 SFM 6 1 Slurry Flow Meter FLOW METER DATA SHEET ...
Page 41: ......