Page 12
SFM 6.1 Slurry Flow Meter
CALIBRATION
Press the
to
Calibration
and
to enter. Use
or
to position
before
each menu item and
to enter. When settings are completed press
to store and
return to the Calibration menu.
20mA at
(
5V Flo
)
Press
then
or
to change the numbers and decimal point.
Use this menu to set the corresponding flow rate that will be
represented by 20mA analog output. If maximum flow is
unknown, enter an estimated flow rate and observe actual flow
to determine the correct maximum value. Any velocity or flow
rate up to +40 ft/sec (12.2 m/sec) may be selected.
4mA at
(
0V Flo
)
Press
or
to set the flow rate corresponding to 4mA analog
output. This setting may be left at zero flow (or velocity or can
be raised to any value less than the 20mA setting, or lowered to
any velocity or corresponding flow rate down to -40 ft/sec
(-12.2 m/sec).
Min Vel
Press
and enter a minimum velocity cutoff. Forward and
reverse velocities less than
Min Vel
will be forced to zero.
Pipe ID
Place the cursor under the digits and then
or
to change the
numbers and decimal point.
Pipe ID
should be entered as the
exact inside diameter of the pipe where the sensor is mounted.
Refer to the Pipe Charts Appendix in this manual for inside
diameter of common pipe types and sizes.
Damping
Increase damping to stabilize readings under turbulent flow
conditions. Decrease for fast response to small changes in flow.
Damping is shown in percentage (maximum is 99%). Factory
default is 10%.
Press
to return to Menu Selections screen.
Summary of Contents for SFM 6.1
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 SFM 6 1 Slurry Flow Meter CONNECTIONS ...
Page 7: ...Page 7 SFM 6 1 Slurry Flow Meter CALIBRATION MENU ...
Page 8: ...Page 8 SFM 6 1 Slurry Flow Meter ICONS ...
Page 19: ...Page 19 SFM 6 1 Slurry Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS ...
Page 28: ...Page 28 SFM 6 1 Slurry Flow Meter FLOW METER DATA SHEET ...
Page 41: ......