GREISINGER GMH 5530 Operating Manual Download Page 13

H80.0.02.6C-16 

 Operating Manual GMH 5530 

page 13 of 20 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

Menu 

Parameter 

Value 

Description

 

 

 

 

 

 or 

 

 

 

 

 

 

Universal Output

 

 

 

oFF 

Interface and analog output off -> minimal power consumption 

SEr:  

Serial interface activated  

dAC:  

Analog output activated 

 

01,11..91 

Base address for serial interface communication 

 

 

 

Set Corr: Input adjustment 

 

 

 

Zero adjustment / offset of voltage measurement 

 

 

oFF 

No zero adjustment for voltage measurement 

-

10.0 … 10.0 

mV 

Offset of voltage measurement in mV 

 

Slope adjustment of voltage measurement 

 

 

oFF 

No slope adjustment for voltage measurement 

-

5.000 … 

5.000 % 

Slope correction of voltage measurement in % 

 

Zero adjustment / offset of temperature measurement 

 

 

oFF 

No zero adjustment for temperature measurement 

-

5.0 … 5.0°C 

Offset of temperature measurement in °C 

 

Slope adjustment of temperature measurement 

 

 

oFF 

No slope adjustment for temperature measurement 

-

5.00 … 5.00% 

Slope correction of temperature measurement in % 

 

Set Clock: Settings for real time clock 

 

 

 

HH:MM 

Clock

: set time 

hours:minutes 

 

 

 

YYYY 

Year

:   set year 

 

 

 

TT.MM 

Date

:   set date 

day.month 

 

 

 

rEAd CAL: Read calibration data:  
p.r.t. chapter 10.2 

“Calibration storage (rEAd CAL)”

 

 

 

8  Output / External Supply 

The device has a serial output (for interface converter USB 5100). If the output is not needed, it is strongly 
recommended to deactivate it (Out oFF) to lower power consumption. This increases battery life time.  
If the device is used together with interface adapter USB 5100 the device is supplied from the interface. 

Pin assignment: 

 

 
4: external 5V, 50mA 
3: GND 
2: TxD/RxD (3.3V Logic) 
1: not used 
 

 

 

 

 

Only suitable adaptor cables are 
permitted (accessories)!

 

The device can be connected to a USB interface of a PC by the electrically isolated interface converter 
USB 5100 (accessory). The data is transmitted binary-coded and protected against transmission errors by 
complex safety mechanism (CRC).  

The following standard software packages are available: 

 

GSOFT3050

:  

Operating and evaluation software for the integrated logger function 

 

EBS20M / -60M

20-/60-channel software for measuring value display 

 

GMHKonfig

Configuration Software (for free on internet) 

In case you want to develop your own software we offer a 

GMH3000-development package 

including: 

  a universally applicable Windows functions library ('GMH3x32e.DLL') with documentation, can be used 

by all ‘established’ programming languages, suitable for:

  

Windows XP™, Windows Vista™, Windows 7™ 

  Programming examples

 

Visual Basic

™, Delphi 1.0™, Testpoint™ etc. 

The device has 2 channels: 

channel 1: actual-value-channel pH, mV or rH and base address 

channel 2: temperature value 

Summary of Contents for GMH 5530

Page 1: ...lly read these instructions before use Please consider the safety instructions Please keep for future reference PCE Instruments France EURL 76 Rue de la Plaine des Bouchers 67100 Strasbourg France Tel...

Page 2: ...tomatic input of pH value from preceding pH measurement 8 6 4 PH ELECTRODE 9 6 4 1 Design 9 6 4 2 Further Information 9 6 4 3 pH electrode suggestions 9 6 5 CALIBRATION OF PH MEASUREMENT 10 6 5 1 How...

Page 3: ...This enables an automatic temperature compensation ATC for pH rH and mVH measurements and displaying the media s temperature The safety requirements see below have to be observed The device must be us...

Page 4: ...creases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 5530 with 2 AAA batteries Operation manual Short form manu...

Page 5: ...7 stab arrow indicates stable measuring value 8 cal arrow indicates a running calibration at operation mode pH 9 No function 4 2 Pushbuttons On off key backlight press shortly activate backlight or sw...

Page 6: ...cted from outside Separately lead through reference electrode it is connected from inside 4 4 Pop up clip Handling Pull at label open in order to swing open the pop up clip Pull at label open again to...

Page 7: ...equal to the concentration of dissolved hydronium ions log H H mol l with H concentration of dissolved hydronium ions in mol l H activity coefficient normally lower than 1 The abbreviation pH stands...

Page 8: ...measurement For example it is used to describe the anti oxidative effect of food This is a measure for the ability of food to reduce harmful free radicals To measure the rH value of a solution procee...

Page 9: ...agm Examples with solutions containing protein like they are used on the medical and biological sector KCI may result in the denaturation of the protein coagulated varnish solutions with a relatively...

Page 10: ...easuring as well as during calibration Otherwise enter actual buffer temperature as accurate as possible see below When working with the standard or DIN buffer series the influences of buffer temperat...

Page 11: ...d series this is pH 10 0 and stir gently As soon as the measured pH value got stable the next calibration step will be displayed No temperature sensor manual input of temperature of buffer 3 Use or to...

Page 12: ...row pH pH value measurement Resolution pH Resolution of ph display 0 1 0 001 tenth pH thousandth pH Calibration Select number of calibration points 1 Pt 1 point only offset calibration slope 59 2mV pH...

Page 13: ...orage rEAd CAL 8 Output External Supply The device has a serial output for interface converter USB 5100 If the output is not needed it is strongly recommended to deactivate it Out oFF to lower power c...

Page 14: ...he device provides the following functions to help with this The usage of the GLP functions is only reasonable if the electrode is not changed Although all data is stored in the device it refers to th...

Page 15: ...tteries Before changing batteries please read the following instruction and follow it step by step Not following the instruction may cause harm to the instrument or the protection against ingress of w...

Page 16: ...1 point calibration Buffer solution defective Use new buffer solution Electrode defective Clean electrode and calibrate again if error occurs again replace electrode Slope is too low Electrode defecti...

Page 17: ...ated pH calibration Automatic 1 2 or 3 point calibration either DIN 19266 buffer or technical buffer GPH PHL Manual 1 2 or 3 point calibration GLP adjustable calibration intervals 1 to 365 days CAL wa...

Page 18: ...H80 0 02 6C 16 Operating Manual GMH 5530 page 18 of 20 _____________________________________________________ _____________________________________________________________________________...

Page 19: ...A device calibrated only at 1 point assures only accurate measuring values at a range close to the buffer value Buffer capacity The pH value of a buffer solution changes only very little when small a...

Page 20: ...H80 0 02 6C 16 Operating Manual GMH 5530 page 20 of 20 _____________________________________________________ _____________________________________________________________________________...