GREISINGER GMH 5530 Operating Manual Download Page 10

H80.0.02.6C-16 

 Operating Manual GMH 5530 

page 10 of 20 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

6.5  Calibration of pH measurement 

The electrode data of pH electrodes are subject to fluctuation due to ageing and manufacturing tolerances. 
Therefore it is necessary to check the calibration with buffer solutions before measurements take place. If 
deviations are too large, a recalibration is necessary. See also: Chapter 10 GLP 

Buffer solutions are liquids with an accurate pH-value. The following buffers can be used for calibration: 

- Technical buffer series 

PHL

 (ready to use, pH 4.01, pH 7.00 und pH 10.01) 

- Standard series 

GPH

 (buffer capsules to be mixed with water pH 4.01, pH 7.00 and pH 10.01) 

- DIN series 

CAL dIn 

(pH 1.68 (A), pH 4.01 (C), pH 6.87 (D), pH 9.18(F) und pH 12.45(G)) 

- Arbitrary buffer 

CAL Edit 

(neutral buffer ranging from 6.5 ... 7.5pH) 

Service life of a buffer solution is limited and will be further reduced unless the electrodes are 
properly rinsed and dried when changing over the solutions. This may even result in incorrect 
calibration! We recommend to use new buffer solution for calibration, as far as possible, and to 
rinse with deionized or distilled water.  

6.5.1  How to prepare calibration buffers of standard GPH series (capsules) 

  

See Notes B.  

6.5.2  Automatic temperature compensation during calibration 

Both the signal of the pH-electrode and the pH-buffer are depending on temperature. If a temperature probe 
is connected, the temperature influence of the electrode is compensated automatically during measuring as 
well as during calibration. Otherwise enter actual buffer temperature as accurate as possible (see below).  
When working with the standard or DIN-buffer series, the influences of buffer temperature are also 
compensated. If buffers are entered manually, make sure to enter the pH-values of the buffers at the relevant 
temperature to ensure optimum calibration of the device.  

6.5.3  How to carry out calibration 

Please note: the calibration can only carried out at a temperature range of 0 - 60°C !

  

If you have not yet done so set device to measuring mode ’pH’. Make sure that either the 

1-, 2- or 3- point 

calibration

 (whichever is required) and desired buffer series (

PHL, GPH, dIn

 or 

Edit

) the has been activated 

(further information in chapter 7 

“Configuration”).

 

Carefully remove electrode safety cap (Attention! Contains 3 mol KCl!).

 

Rinse electrode with distilled water and dry it carefully. 

How to start calibration: press 

 key for 2 seconds.

  

The display will prompt you to measure the first calibration solution.  

You can abort calibration at any time by pressing 

 key. In such a case the last calibration before this one 

remains valid. 

1. Calibration point 1: 

‘Pt. 1’ 

 

 

*1) 

Place electrode and temperature probe 
(if any) in the neutral solution stirring 
gently.  

(For 1-point calibration: solutions with 
arbitrary pH value (e.g. pH 4) can be uses) 

As soon as the measured pH value got 
stable the next calibration step will be 
displayed.  

 

No temperature sensor:  
manual input of  
temperature of buffer 1 

Use  

 or 

 to enter the temperature of 

the buffer solution.  

Use 

 to confirm the value; the next 

calibration step is displayed. 

If 1-point calibration is chosen the calibration is already done at this point, the bar graph display on the left 
shows the electrode’s state rating.   

2. Rinse electrode in distilled or deionized water, dry electrode 

 

 

Summary of Contents for GMH 5530

Page 1: ...lly read these instructions before use Please consider the safety instructions Please keep for future reference PCE Instruments France EURL 76 Rue de la Plaine des Bouchers 67100 Strasbourg France Tel...

Page 2: ...tomatic input of pH value from preceding pH measurement 8 6 4 PH ELECTRODE 9 6 4 1 Design 9 6 4 2 Further Information 9 6 4 3 pH electrode suggestions 9 6 5 CALIBRATION OF PH MEASUREMENT 10 6 5 1 How...

Page 3: ...This enables an automatic temperature compensation ATC for pH rH and mVH measurements and displaying the media s temperature The safety requirements see below have to be observed The device must be us...

Page 4: ...creases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 5530 with 2 AAA batteries Operation manual Short form manu...

Page 5: ...7 stab arrow indicates stable measuring value 8 cal arrow indicates a running calibration at operation mode pH 9 No function 4 2 Pushbuttons On off key backlight press shortly activate backlight or sw...

Page 6: ...cted from outside Separately lead through reference electrode it is connected from inside 4 4 Pop up clip Handling Pull at label open in order to swing open the pop up clip Pull at label open again to...

Page 7: ...equal to the concentration of dissolved hydronium ions log H H mol l with H concentration of dissolved hydronium ions in mol l H activity coefficient normally lower than 1 The abbreviation pH stands...

Page 8: ...measurement For example it is used to describe the anti oxidative effect of food This is a measure for the ability of food to reduce harmful free radicals To measure the rH value of a solution procee...

Page 9: ...agm Examples with solutions containing protein like they are used on the medical and biological sector KCI may result in the denaturation of the protein coagulated varnish solutions with a relatively...

Page 10: ...easuring as well as during calibration Otherwise enter actual buffer temperature as accurate as possible see below When working with the standard or DIN buffer series the influences of buffer temperat...

Page 11: ...d series this is pH 10 0 and stir gently As soon as the measured pH value got stable the next calibration step will be displayed No temperature sensor manual input of temperature of buffer 3 Use or to...

Page 12: ...row pH pH value measurement Resolution pH Resolution of ph display 0 1 0 001 tenth pH thousandth pH Calibration Select number of calibration points 1 Pt 1 point only offset calibration slope 59 2mV pH...

Page 13: ...orage rEAd CAL 8 Output External Supply The device has a serial output for interface converter USB 5100 If the output is not needed it is strongly recommended to deactivate it Out oFF to lower power c...

Page 14: ...he device provides the following functions to help with this The usage of the GLP functions is only reasonable if the electrode is not changed Although all data is stored in the device it refers to th...

Page 15: ...tteries Before changing batteries please read the following instruction and follow it step by step Not following the instruction may cause harm to the instrument or the protection against ingress of w...

Page 16: ...1 point calibration Buffer solution defective Use new buffer solution Electrode defective Clean electrode and calibrate again if error occurs again replace electrode Slope is too low Electrode defecti...

Page 17: ...ated pH calibration Automatic 1 2 or 3 point calibration either DIN 19266 buffer or technical buffer GPH PHL Manual 1 2 or 3 point calibration GLP adjustable calibration intervals 1 to 365 days CAL wa...

Page 18: ...H80 0 02 6C 16 Operating Manual GMH 5530 page 18 of 20 _____________________________________________________ _____________________________________________________________________________...

Page 19: ...A device calibrated only at 1 point assures only accurate measuring values at a range close to the buffer value Buffer capacity The pH value of a buffer solution changes only very little when small a...

Page 20: ...H80 0 02 6C 16 Operating Manual GMH 5530 page 20 of 20 _____________________________________________________ _____________________________________________________________________________...

Reviews: