background image

H33.0.02.6C-09 

Operating Manual GMH 3530 

page 3 of 14

 

 

Operating and Maintenance Advice: 

a)  When to replace battery: 

If 

 and ´bAt´ are shown in the lower display the battery has been used up and needs to be replaced. The de-

vice will, however, operate correctly for a certain time. 
If ´bAt´ is shown in the upper display the voltage is too low to operate the device; the battery has been completely 
used up. 

Please note:  The battery has to be taken out, when storing device above 50°C. 

We recommend to take out battery if device is not used for a longer period of time. 

b)  Treat device and sensor carefully. Use only in accordance with above specification. (do not throw, hit against etc.). 

Protect plug and socket from soiling. 

c)  When connecting the temperature probe the connector may not lock to the jack correctly. In such a case hold the 

connector not at the case but at the buckling protection of the cable during the plug in. Don’t connect electrode 
canted! If plug is entered correctly, it will slide in smoothly. 

 

To disconnect temperature probe do not pull at the cable but at the plug. 

 

If plug is entered incorrectly the connecting pins of the plug can be damaged. => Plug can no longer be used and 
connecting cable needs to be replaced. 

d)  Mains operation: 

When using a power supply device please note that operating voltage has to be 10.5 to 12 V DC. 
Do not apply overvoltage!! Cheap 12V-power supply devices often have excessive no-load voltage. We, there-
fore, recommend using regulated voltage power supply devices. Trouble-free operation is guaranteed by our 
power supply devices. Trouble-free operation is guaranteed by our power supply, GNG10/3000.  
Prior to connecting the plug power supply device with the mains supply make sure that the operating voltage 
stated at the power supply device is identical to the mains voltage. 

e)  Display values for damaged electrode cable or if no pH or redox-electrode has been connected 

If no electrode is connected or the connection cable is damaged the display will nevertheless show mV, pH or rH 
values. Please note that these values can never be correct measuring results! 

Safety Requirements: 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and 
special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any 

other climatic conditions than those stated under "Specification". 

2.  If the device is transported from a cold to a warm environment condensation may cause in a failure of the func-

tion. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a 
new start-up. 

3.  If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed most care-

fully. Internal connection in third party devices (e.g. connection GND and earth) may result in not-permissible vol-
tages impairing or destroying the device or another device connected. 

 

Warning: 

If device is operated with a defective mains power supply (short circuit from mains voltage to output 

voltage) this may result in hazardous voltages at the device (e.g. sensor socket, serial interface). 

4.  If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be 

marked accordingly to avoid re-starting. 

 

Operator safety may be a risk if: 

 

 - there is visible damage to the device 

 

 - the device is not working as specified 

 

 - the device has been stored under unsuitable conditions for a longer time. 

 

In case of doubt, please return device to manufacturer for repair or maintenance. 

5. 

Warning: 

Do not use these products as safety or emergency stop devices or in any other application where failure 

of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material damage. 

 

Summary of Contents for GMH 3530

Page 1: ...09 GREISINGER electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for pH Redox temperature as of version V2 6 Operating Manu...

Page 2: ...AND SYSTEM MESSAGES 10 13 THE PH ELECTRODE 11 13 1 VARIOUS APPLICATIONS REQUIRE SPECIAL ELECTRODES 11 14 THE REDOX ELECTRODE 12 15 THE SERIAL INTERFACE 13 15 1 THE FOLLOWING INTERFACE FUNCTIONS WILL B...

Page 3: ...s connected or the connection cable is damaged the display will nevertheless show mV pH or rH values Please note that these values can never be correct measuring results 4 Safety Requirements This dev...

Page 4: ...ed press for 2 sec the min or max value will be deleted Configuration to enter values or change settings CAL for pH mode only press shortly display state of electrode condition and calibration data pr...

Page 5: ...for pH measurements only 2 Pt the pH electrode will be calibrated at 2 points one calibration point in neutral range and one calibration point in acid or basic range 3 Pt the pH electrode will be cal...

Page 6: ...measuring mode 8 Manual Setting of pH Value for measuring function rH Running the GMH 3530 in the rH measuring mode requires both temperature and pH value Press shortly again to switch over to the pH...

Page 7: ...mum calibration of the device 9 3 How to carry out calibration Please note the calibration can only carried out at a temperature range of 0 60 C If you have not yet done so set device to measuring mod...

Page 8: ...of temperature for solution 2 Use or keys to enter the buffer solution temperature Use to take over value and to display electrode state 5 Rinse electrode in distilled or non ionised water dry electro...

Page 9: ...ermine the rH value of your solution proceed as follows Please note Make sure that pH and redox electrodes are in a perfect condition during measuring and that they are cleaned and dried carefully bef...

Page 10: ...eck are there any values below the measur ing range specified meas device not suitable Electrode sensor cable defective replace electrode probe System fault switch on again if fault continues to exist...

Page 11: ...embrane or the diaphragm will influence the measurements and have to be removed at regular intervals This can be done by means of automatic cleaning equipment 13 1 Various applications require special...

Page 12: ...mometer Check reference electrolyte level at regular intervals if necessary top up electrolyte level with 3 mol lKCI solution through filling hole closed by means of a silicon ring using a syringe or...

Page 13: ...ting mode pH mV rH Ther Channel 1 2 1 2 1 2 1 x x x x x x x 0 Read nominal value x x x 1 Set nominal value x x x x x x x 3 Read system status x x x x x x x 6 Read min value x x x x x x x 7 Read max va...

Page 14: ...elay can be set to values between 1and 120 min it can be completely deactivated Housing dimensions 142 x 71 x 26 mm L x W x D impact resistant ABS plastic housing membrane key board transparent panel...

Reviews: