2 Configuration
To change device settings, press
Menu
(key 4) for 2 seconds. This will call the configuration menu (main
display: „SEt“).
Pressing key
Menu
changes between the menus, pressing
4
(key 3) jumps to the referring parameters,
which can be selected with key
4
(key 3).
The parameters can be changed with
5
(key 2) or
6
(key 5).
Pressing
Menu
again jumps back to the main configuration menu and saves the settings.
Quit
(key 6) finishes the configuration and returns to standard measuring operation.
Menu
Param.
Values
Meaning
‚Menu‘
4
4
4
4
5
or
6
C_l
og
C_dat
Sto
r
Set Configuration: Generic Configurations
Unit
mbar,bar..
Unit
: Unit of display
*
*
Rate
: Measuring rate (p.r.t. chapter 2.1)
*
*
Slo
Slow
measuring rate (4Hz filtered, low power consumption)
*
*
FASt
Fast
measuring rate, filtered (>1000Hz)
*
*
rAtE
P.dEt
Peak detection
: fast measuring rate, unfiltered (>1000Hz)
*
*
1-120
Averaging period in seconds, used by the averaging function
*
*
t.AVG
oFF
Averaging function deactivated
*
*
1-120
Auto Power Off
time in minutes
P.oFF
oFF
Auto Power Off
deactivated
oFF
Function of the output: No output function, lowest power consumption
SEr
Output is serial interface
Out
dAC
Output is analogue output 0...1V
Adr.
01,11..91
Base
address
of interface
dAC.0
-100 ... 2000
mbar
Enter desired value at which the analogue output potential should be 0V
(if Out = dAC)
SEt
ConF
dAC.1
-100 ... 2000
mbar
Enter desired value at which the analogue output potential should be 1V
(if Out = dAC)
Set Calibration: Adjustment of Sensor
*
-500 ... 500
mbar
The
offset of sensor
will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
*
OFFS
oFF
Zero displacement inactive (=0.00)
*
-2.000 ...
2.000
The measuring
scale of sensor
will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device
*
SEt
CAL
SCAL
oFF
Scale correction factor inactive (=0.000)
*
Set Alarm: Settings Of Alarm Function
on
Alarm
on, with horn-sound
no.So
Alarm on, without horn-sound
AL.
oFF
Alarm deactivated
AL.Lo
-100 mbar ...
AL.Hi
Min alarm rail (not when AL. oFF, Sensor-Min is the lower display range of
connected sensor)
SEt
AL.
AL.Hi
AL.Lo ...
2000 mbar
Max alarm rail (not when AL. oFF, Sensor-Max is the upper display range
of connected sensor)
Set Logger: Configuration Of Logger Function
*
*
CYCL
Cyclic
: logger function ‚cyclic logger‘
* * *
Stor
Store
: logger function ‚individual value logger‘
* * *
Func
oFF
no logger function
* * *
CYCL
1..3600
Cycle time of cyclic logger [seconds]
*
*
SEt
LoGG
Lo.Po
on/oFF
Low-power logger
with very low power consumption
(only for cyclic logger and slow measuring rate)
*
*
Set Clock: Setting Of Real Time Clock
CLOC
HH:MM
Clock
: Setting of time hours:minutes
dAtE
TT.MM
Date
: Setting of date
day.month
SEt
CLOC
YEAr
YYYY
Year
: Setting of year
C_log: cyclic logger (active)
C_dat: cyclic logger with data (no active logging)
C_stor: individual value logger with data
Note:
When using the logger function some settings in the menu may not be accessible (*).
If this settings should be changed, the logger has to be stopped before, eventually the logger data
has to be cleared. (p.r.t. chapter 3)