background image

H66.0.01.6C-10 

Operating Manual GLF 100 

Page 5 of 8 

 

 

 

8  Configuration of the instrument 

To configure the instruments according to Your needs proceed like follows: 

  Switch off the instrument. 

  Switch on the instrument and press mode-key (

)

 during the segment test 

(

 )

 until 

the dis

play shows the first parameter “P_oF” 

  If a parameter should be edited, press key up or down ( 

 

 ), the setting of the 

parameter will be shown.  

You can be changed via up/down keys and receipt the changed parameter with 

  Jump to the next parameter with 

 

Parameter 

Value 

Meaning

 

Key 

 

 

Key 

 

 

 

 

Auto Power-Off

   

factory setting: 20 min.

 

 

Auto Power-Off

  in minutes. 

If there won’t be pressed any key for the time of 

the power off time setting, the device will be switched off automatically to 
save battery power (adjustable range 1 to 120 min) 

 

automatic power-off is deactivated (continuous operating) 

 

Unit and range of display

   

factory setting: Auto

 

Auto 
µS/cm+mS/cm 

Measuring of conductivity with auto-range 

100.0 mS/cm 

Measuring of conductivity, range 0.0 ... 100.0 mS/cm (no auto-range) 

20.00 mS/cm 

Measuring of conductivity, range 0.00 ... 20.00 mS/cm (no auto-range) 

2000 µS/cm 

Measuring of conductivity, range 0 ... 2000 µS/cm (no auto-range) 

2000 mg/l 

Measuring of TDS, 0 ... 2000 mg/l 

50.0 SAL 

Measuring of Salinity, 0.0 .. 50.0 g/kg 

 

Unit of temperature display

  

factory setting: °C

 

°C 

temperature is displayed in °C (factory setting) 

°F 

temperature is displayed in °F 

 

Temperature compensation 

(only for µS/cm and mS/cm, p.r.t. note below)

  

factory setting: nLF

 

oFF 

No temperature compensation of conductivity measuring 

nLF 

non-linear temperature compensation for natural water acc. to EN 27888 
(ISO 7888) to measure ground water, surface water, drinking water and 
purest water (factory setting) 

 

Reference temperature of temp. compensation 

(only for t.Cor <> oFF)

   factory setting: 25 °C

 

25 °C / 77 °F 

Reference temperature 25 °C / 77 °F 

20 °C / 68 °F 

Reference temperature 20 °C / 68 °F 

 

Setting of TDS-factor 

  

factory setting: 0.40

 

0.40...1.00 

Setting of factor of the TDS measurement 

 

Restore to factory settings

 

 

Settings are kept 

 

Restoring device to factory settings 

Pressing 

 again stores the settings, the instruments restarts (segment test) 

Please note:   If there is no key pressed within the menu mode within 2 minutes, the 

configuration will be cancelled, the entered settings are lost! 

 

Note for temperature compensation: To measure the TDS value always use non-linear 

temperature compensation for natural water (reference temperature = 25 °C). 
To measure salinity the instrument automatically switches over to the non-linear 
temperature compensation acc. to IOT (reference temperature = 15 °C).. 

 

 

Summary of Contents for GLF 100 RW

Page 1: ...chnik GmbH Standort Greisinger Hans Sachs Str 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de WEEE Reg Nr DE 93889386 Conductivity Measuring Instrument as of version 1 2 Op...

Page 2: ...shbuttons 4 7 1 Min max value memory 4 7 2 Hold function 4 8 Configuration oft he instrument 5 9 Adjusting oft the instrument 6 10 System Messages 6 11 Inspection of the accuracy Adjustment Services 6...

Page 3: ...result in not permissible voltages impairing or destroying the device or another device connected 4 Whenever there may be a risk whatsoever involved in running it the device has to be switched off im...

Page 4: ...reset Min and Max value hold Hold current measuring value HLD in display 7 1 Min max value memory The lowest and highest measured value since the turning on of the device will be stored The stored va...

Page 5: ...easuring of TDS 0 2000 mg l 50 0 SAL Measuring of Salinity 0 0 50 0 g kg Unit of temperature display factory setting C C temperature is displayed in C factory setting F temperature is displayed in F T...

Page 6: ...setting of temperature measurement see below Scale correction of temperature 5 00 5 00 Scale adjustment of temperature measurement Display measured value offset 1 scale adjustment 100 Pressing again s...

Page 7: ...f the instrument is very constant if it is used as intended Depending on the necessary accuracy the instruments can be used up to several years without recalibration of the cell constant If the accura...

Page 8: ...instrument 25 50 C Temperature cell 5 80 C short time 100 C Relative humidity instrument 0 95 RH non condensing Storage temperature 5 50 C Housing Dimension 110 x 67 x 30 mm H x W x D impact resistant...

Reviews: