12 Adjusting of the instrument 

To adjust the instruments according to your needs proceed like follows: 

  Switch instrument off. 
   Press 

zero

 button and keep it pressed. Turn the device on. Release the mode button after the 

segment test (             ) finished. The first parameter “

SCL

“ is displayed. 

 

  If a parameter should be changed press key up or down (                   ) the present setting of the 

parameter will be shown and can be changed via up/down keys.  

The parameter setting is confirmed by pressing 

 . 

  Jump to the next parameter with 

 . 

 

Parameter  value 

information

 

button

 

 

buttons

 

 

Offset of sensors   

factory setting

: off

 

OFF 

-5.0 … +5.0

 

 Selectable -5.0 … +5.0 

Scale correction    

factory setting

: off = 0%

 

OFF 

-5.00 … +5.00

 

Adjustment in % 

Pressing  

 again stores the settings, the instruments restarts (segment test) 

Please note:   

If there is no key pressed within the menu mode within 2 minutes, the 
adjusting will be cancelled, the entered settings are lost! 

 

13 Notes to the calibration services 

Calibration certificates - DKD-certificates - other certificates: 

If device should be certificated for its accuracy, it is the best solution to return it with the referring sensors to 

the manufacturer. (please state desired test values, e.g. 1bar abs) 

Only the manufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy!

 

14 Error and system messages 

ERR.  1 

measuring range has been exceeded 

ERR. 2 

measured value has fallen below permitted range 

ERR. 3 

display range has been exceeded (>19999) 

ERR. 4 

measuring value has fallen below displayable range (< -1999) 

ERR. 7 

System fault - the device has detected a system fault (defective or far outside allowable 

ambient temperature range) 

--- 

Sensor error or value could not be calculated 

 

If the symbol “

LOBAT

” is displayed at the left side of display, the battery is weak, measuring can be 

continued for a short time. 
If “

BAT

 is displayed in the display the battery is used up and needs to be replaced. Measuring is no 

more possible. 

 

Summary of Contents for GDH 200-14

Page 1: ...H55 0 31 6C 05 from Version 1 1 GDH 200 14 Operating Manual Vacuum Baro Manometer...

Page 2: ...3 SAFETY INSTRUCTIONS 3 4 OPERATION NOTES 3 5 OPERATION 4 5 1 DISPLAY ELEMENTS 4 5 2 PUSHBUTTONS 4 6 STARTING 4 7 PRESSURE CONNECTION 4 8 MIN MAX VALUE MEMORY 5 9 ZERO FUNCTION RELATIVE MEASURING 5 10...

Page 3: ...and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using it 1 Trouble free operation and reliability of th...

Page 4: ...tivates deactivates the Zero function 6 Starting Switch the device on with button After the segment test the device displays some information to its current configuration OFFS if offset correction is...

Page 5: ...off 9 Zero function relative measuring With help of the Zero function relative measurements can be realized press zero button for 2s NULL is displayed shortly and the displayed value is set to 0 Press...

Page 6: ...20 min 1 120 Auto Power Off turn off delay in minutes If no key is pressed for the time adjusted in this parameter the device is automatically switched off to safe battery power adjustable range 1 to...

Page 7: ...f there is no key pressed within the menu mode within 2 minutes the adjusting will be cancelled the entered settings are lost 13 Notes to the calibration services Calibration certificates DKD certific...

Page 8: ...ue is set to zero Display approx 13 mm high 4 digit LCD display Pushbuttons 3 membrane keys for on off min max value request offset adjustment Working conditions 25 to 50 C 0 to 80 RH non condensing S...

Reviews: