H55.0.31.6C-05
Operating Manual GDH 200-14
page 7 of 8
_____________________________________________________
_____________________________________________________________________________
12 Adjusting of the instrument
To adjust the instruments according to your needs proceed like follows:
Switch instrument off.
Press
zero
button and keep it pressed. Turn the device on. Release the mode button after the
segment test ( ) finished. The first parameter “
SCL
“ is displayed.
If a parameter should be changed press key up or down ( ) the present setting of the
parameter will be shown and can be changed via up/down keys.
The parameter setting is confirmed by pressing
.
Jump to the next parameter with
.
Parameter value
information
button
buttons
Offset of sensors
factory setting
: off
OFF
-5.0 … +5.0
Selectable -5.0 … +5.0
Scale correction
factory setting
: off = 0%
OFF
-5.00 … +5.00
Adjustment in %
Pressing
again stores the settings, the instruments restarts (segment test)
Please note:
If there is no key pressed within the menu mode within 2 minutes, the
adjusting will be cancelled, the entered settings are lost!
13 Notes to the calibration services
Calibration certificates - DKD-certificates - other certificates:
If device should be certificated for its accuracy, it is the best solution to return it with the referring sensors to
the manufacturer. (please state desired test values, e.g. 1bar abs)
Only the manufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy!
14 Error and system messages
ERR. 1
measuring range has been exceeded
ERR. 2
measured value has fallen below permitted range
ERR. 3
display range has been exceeded (>19999)
ERR. 4
measuring value has fallen below displayable range (< -1999)
ERR. 7
System fault - the device has detected a system fault (defective or far outside allowable
ambient temperature range)
---
Sensor error or value could not be calculated
If the symbol “
LOBAT
” is displayed at the left side of display, the battery is weak, measuring can be
continued for a short time.
If “
BAT
”
is displayed in the display the battery is used up and needs to be replaced. Measuring is no
more possible.