H55.0.31.6C-05 

Operating Manual GDH 200-14  

page 6 of 8 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

11 Configuration of the device 

Follow this instructions to configure the functions of the device: 

  Switch the device off. 
   Press 

mode

 button and keep it pressed. Turn the device on. Release the mode button after the 

segment test (             ) finished. The first parameter “

P.OFF

“ is displayed. 

 

  If a parameter should be changed press key up or down (                  ) the present setting of the 

parameter will be shown and can be changed via up/down keys.  

The parameter setting is confirmed by pressing 

 . 

  Jump to the next parameter with 

 . 

Parameter 

value 

information

 

button

 

 

buttons

  

 

P.OFF 

Auto Power-Off

 (turn-off delay)  

factory setting: 20 min.

 

 120 

Auto Power-Off

 (turn-off delay) in minutes. If no key is pressed for the 

time adjusted in this parameter, the device is automatically switched off to 

safe battery power. (adjustable range 1 to 120 min) 

OFF 

automatic power-off is deactivated (continuous operating) 

UNIT 

Unit and range of display

   

factory setting: nbar=mbar

 

PSI 

0 … 160.00 PSI, resolution 0.02 PSI 

nnHG 

0 … 8250 mmHg, resolution 1mmHg 

nbAr 

0 … 11000 mbar, resolution 1 mbar 

bAr 

0 … 11.000 bar, resolution 0.001 bar 

Sea-level correction

    

factory setting

: off

 

OFF 

Sea-level correction off 

ON 

Sea-level correction on 

ALTI 

Altitude input for sea-level correction

    

factory setting

: 340

 

-500 … 9000 

-500 … 9000 m selectable 

INIT

 

Restore to factory settings

 

NO

 

Settings are kept 

YES 

Restoring device to factory settings 

Pressing 

 again stores the settings, the instruments restarts (segment test) 

Please note: 

If there is no key pressed within the menu mode within 2 minutes, the 
configuration will be cancelled, the entered settings are lost! 

 

Summary of Contents for GDH 200-14

Page 1: ...H55 0 31 6C 05 from Version 1 1 GDH 200 14 Operating Manual Vacuum Baro Manometer...

Page 2: ...3 SAFETY INSTRUCTIONS 3 4 OPERATION NOTES 3 5 OPERATION 4 5 1 DISPLAY ELEMENTS 4 5 2 PUSHBUTTONS 4 6 STARTING 4 7 PRESSURE CONNECTION 4 8 MIN MAX VALUE MEMORY 5 9 ZERO FUNCTION RELATIVE MEASURING 5 10...

Page 3: ...and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using it 1 Trouble free operation and reliability of th...

Page 4: ...tivates deactivates the Zero function 6 Starting Switch the device on with button After the segment test the device displays some information to its current configuration OFFS if offset correction is...

Page 5: ...off 9 Zero function relative measuring With help of the Zero function relative measurements can be realized press zero button for 2s NULL is displayed shortly and the displayed value is set to 0 Press...

Page 6: ...20 min 1 120 Auto Power Off turn off delay in minutes If no key is pressed for the time adjusted in this parameter the device is automatically switched off to safe battery power adjustable range 1 to...

Page 7: ...f there is no key pressed within the menu mode within 2 minutes the adjusting will be cancelled the entered settings are lost 13 Notes to the calibration services Calibration certificates DKD certific...

Page 8: ...ue is set to zero Display approx 13 mm high 4 digit LCD display Pushbuttons 3 membrane keys for on off min max value request offset adjustment Working conditions 25 to 50 C 0 to 80 RH non condensing S...

Reviews: