11 Configuration of the device
Follow this instructions to configure the functions of the device:
Switch the device off.
Press
mode
button and keep it pressed. Turn the device on. Release the mode button after the
segment test ( ) finished. The first parameter “
P.OFF
“ is displayed.
If a parameter should be changed press key up or down ( ) the present setting of the
parameter will be shown and can be changed via up/down keys.
The parameter setting is confirmed by pressing
.
Jump to the next parameter with
.
Parameter
value
information
button
buttons
P.OFF
Auto Power-Off
(turn-off delay)
factory setting: 20 min.
1
…
120
Auto Power-Off
(turn-off delay) in minutes. If no key is pressed for the
time adjusted in this parameter, the device is automatically switched off to
safe battery power. (adjustable range 1 to 120 min)
OFF
automatic power-off is deactivated (continuous operating)
UNIT
Unit and range of display
factory setting: nbar=mbar
PSI
0 … 160.00 PSI, resolution 0.02 PSI
nnHG
0 … 8250 mmHg, resolution 1mmHg
nbAr
0 … 11000 mbar, resolution 1 mbar
bAr
0 … 11.000 bar, resolution 0.001 bar
Sea-level correction
factory setting
: off
OFF
Sea-level correction off
ON
Sea-level correction on
ALTI
Altitude input for sea-level correction
factory setting
: 340
-500 … 9000
-500 … 9000 m selectable
INIT
Restore to factory settings
NO
Settings are kept
YES
Restoring device to factory settings
Pressing
again stores the settings, the instruments restarts (segment test)
Please note:
If there is no key pressed within the menu mode within 2 minutes, the
configuration will be cancelled, the entered settings are lost!