T14.0.1X.6C-04

Operating Manual GLMU 400 MP

page 6 of 8

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             

10 Configuration

In the configuration the devices parameters can be changed.
The jumper has to be set, p.r.t. figure right hand side. To set or
remove jumper, the housing cover has to be removed. Ex works
the jumper is set.
To change parameters press “SET” (key 1) two seconds, then the
parameter selection is started with the first parameter (display
shows “CELL”).
By pressing “SET” the desired parameter is selected, the editing of
the parameter values happens via keys 

5

 (key 3) or 

6

 (key 2).

Pressing “SET” again (after editing the parameter values) returns
to the parameter selection
Pressing “SET” again after the last parameter finishes the configu-
ration, stores the changes and restarts the device.

10.1  'CELL': Setting of the cell-constant of the connected measuring cell

Possible inputs: 0.300...1.200. Unit is 1/cm.
The value was evaluated and stored ex works for the connected measuring cell (in between 0.45...0.65).

10.2  'Func': Setting of the measuring function

Possible inputs:

cond  conductivity
reSi:

resistivity

tdS:

TDS (total dissolved solids

SAL: Salinity = salt content of sea water

10.3  'rAnG': Setting of the measuring range (not at “Func SAL”)

Choice the desired measuring range (range is depending on the function set above).
Possible inputs:

Range (rAnG) r 4

Range (rAnG) r 3 Range (rAnG) r 2 Range (rAnG) r 1

Range (rAnG) r 0

Conductivity (cond) 0 - 500 mS/cm

0,0 - 200,0 mS/cm

0,00 - 20,00 mS/cm 0 - 2000 µS/cm

0,0 - 200,0 µS/cm

Resistivity (rESi)

1,00 - 50,00 Ohm·cm 1,0 - 500,0 Ohm·cm 1 - 5000 Ohm·cm

0,00 - 20,00 kOhm·cm 0 - 200 kOhm·cm

TDS (tdS)

0 - 1000 g/l

0.0 - 200,0 g/l

0.00 - 20,00 g/l

0 - 2000 mg/l

0,0 - 200,0 mg/l

Salinity (SAL)

0,0 - 70,0

10.4  't.Cor': Setting of temperature compensation method (not at “Func SAL”)

Possible inputs: off:  no temperature compensation

nLF:non linear compensation for natural water according to EN 27888 (DIN 38404).

 

For measurings of ground and surface water, drinking water or pure water.

 

Attention: restricted temperature range: -5..105 °C

 

Lin: linear temp. compensation (for other aqueous solutions, factor has to be entered)

10.5  't.Lin': Setting of temperature coefficient (only for t.Cor = Lin)

Possible input: 0.300 .. 5.000 [%]
If it is only measured in a narrow conductivity range, the temperature compensation factor of this range can
be determined.

10.6  't.rEF': Setting of reference temperature (only for t.Cor = nLF or Lin)

Possible input: 20 °C or 25 °C

10.7  'C.tdS': Setting of TDS-factor

Calculation factor for TDS (Total Dissolved Solids)-measurements, possible input: 0.40 .. 1.00.
The calculation factor depends on the composition of the medium and has to be determined
for each type of water.

If the jumper is removed from
the shown contacts, the con-
figuration is inaccessible, val-
ues are protected.

Summary of Contents for GLMU 400 MP

Page 1: ...Conductivity Transmitter with 4 pole measuring cell graphite as of version 2 1 Operating Manual GLMU 400 MP Contents 1 Intended use 2 2 General Advice 2 3 Safety Requirements 2 4 Disposal instructions 2 5 Calibration Service 2 6 Installation 3 7 Operation 4 8 Display functions 5 9 Error and system messages 5 10 Configuration 6 11 Specification 8 ...

Page 2: ...adjusted to the ambient temperature be fore trying a new start up 2 General instructions and safety regulations for electric light and heavy current plants including domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third party devices e g connection GND and earth...

Page 3: ...6 1 Assignment of elbow type plug 4 20 mA 2 wire connection voltage 3 wire connection 1 supply 2 GND signal 1 signal 3 supply voltage Uv 4 supply voltage Uv The type current or voltage output is set by works and cannot be changed 6 2 Terminal assignment of meas cell socket contact no 4 pole cell K 0 55 1 temperature 2 temperature GND 3 supply 1 4 signal 1 5 signal 2 6 supply 2 7 not used 6 3 Usage...

Page 4: ... should be controlled or improved this is done by means of suitable reference solutions and the adjusting of the cell factor Attention Wrong handling of reference solution can make them useless very fast Measuring procedure Especially when measuring low conductivity Before immersion to the measuring solution rinse the measur ing cell with deionised water The measuring is speeded up considerably if...

Page 5: ...CLr After 10 seconds the currently measured values will be displayed again 9 Error and system messages Display Description Possible fault cause Remedy Err 1 measuring range ex ceeded Wrong signal Temperature above 140 C not allowed Choose right measuring range for con ductivity Err 2 Measuring value below measuring range Wrong signal Temperature below 5 C not allowed Choose right measuring range f...

Page 6: ...set above Possible inputs Range rAnG r 4 Range rAnG r 3 Range rAnG r 2 Range rAnG r 1 Range rAnG r 0 Conductivity cond 0 500 mS cm 0 0 200 0 mS cm 0 00 20 00 mS cm 0 2000 µS cm 0 0 200 0 µS cm Resistivity rESi 1 00 50 00 Ohm cm 1 0 500 0 Ohm cm 1 5000 Ohm cm 0 00 20 00 kOhm cm 0 200 kOhm cm TDS tdS 0 1000 g l 0 0 200 0 g l 0 00 20 00 g l 0 2000 mg l 0 0 200 0 mg l Salinity SAL 0 0 70 0 10 4 t Cor ...

Page 7: ...e measuring The offset of the measuring will be shifted by this value the input is in C Calculation see below Max input range 5 0 5 0 C or oFF 0 0 factory setting off 10 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input range 5 00 5 00 or oFF 0 00 factory setting off The adjusting of the temperature...

Page 8: ... mA type 2 wire output signal electrically isolated options AV01 0 1 V AV10 0 10 V 3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20 mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10 mA 0 10 V or refer to type plate Rev voltage protection 50 V permanent Perm impedance at 4 20mA RA Ohm Uv 12 V 0 02 A Example for Uv 18V RA 18 V 12 V 0 02 A RA 300 Ohm Per...

Reviews: