background image

T14.0.1X.6C-04

Operating Manual GLMU 400 MP

page 4 of 8

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             

7 Operation

7.1  General information about the conductivity measuring

7.1.1 The measuring cell

During the measurement, the conductivity measuring cell must be dipped at least in so far, that at least
30 mm beginning from the top of the measuring cell, is located in the medium.
The maximum immersion depth for continuous operation should not exceed 110 mm.

The measuring cell can either be stored dry or in water. After dry storage wetting time will be prolonged
slightly. If changing over from one liquid to another with conductivity varying widely make sure to properly
rinse the measuring cell with deionised water.

Attention:  Measuring cell must never come into contact with water-repellent materials such as oil

or silicone.

If conductivity measured is much higher or lower than expected this may be due to the measuring being
soiled with non-conducting or conducting extraneous materials. The measuring cell has to be cleaned e.g.
with a watery soap solution.
When measuring media at low conductivity the measuring cell has to be stirred sufficiently.

7.1.2 Measuring hints

Conductivity measuring is comparably easy to perform, the precision of the instrument is very constant if it is
used as intended. Depending on the necessary accuracy the instruments can be used up to several years
without re-calibration of the cell constant.
If the accuracy  should be controlled or improved, this is done by means of suitable reference solutions and
the adjusting of the cell factor.
Attention! Wrong handling of reference solution can make them useless very fast.

Measuring procedure:

Especially when measuring low conductivity: Before immersion to the measuring solution, rinse the measur-
ing cell with deionised water.
The measuring is speeded up considerably, if the measuring cell is immersed and pulled out the solution
several times. Especially when measuring low conductivity the measuring cell needs sufficient flow during the
measuring, e.g. via stirring the solution.
Sufficient time has to be waited for, until the measuring cell has adjusted to the measuring solution tem-
perature.

7.1.3 Temperature compensation

The conductivity of aqueous solution is temperature dependent. The dependency itself is strongly dependent
on the kind of solution. For the most applications e.g. in fish farming etc., the non linear temperature com-
pensation of natural waters is precise (“nLF” according to EN 27888). The most common reference tem-
perature is 25 °C.

7.1.4 .Salinity measuring

The salinity (salt content) of seawater can be determined in the measuring mode „SAL“ (basis: International
Oceanographic Tables; IOT)
The salinity of standard-seawater is 35 ‰ ( 35 g salt per 1 kg seawater).
The values are displayed in ‰ (g/kg).

7.1.5 TDS measuring

At the TDS measuring the dry residues or 

t

otal 

d

issolved 

s

olids value is calculated by means of the conduc-

tivity and a calculation factor (C.tdS). This factor has to be evaluated for the referring type of solution.
The values are displayed in mg/l.

Summary of Contents for GLMU 400 MP

Page 1: ...Conductivity Transmitter with 4 pole measuring cell graphite as of version 2 1 Operating Manual GLMU 400 MP Contents 1 Intended use 2 2 General Advice 2 3 Safety Requirements 2 4 Disposal instructions 2 5 Calibration Service 2 6 Installation 3 7 Operation 4 8 Display functions 5 9 Error and system messages 5 10 Configuration 6 11 Specification 8 ...

Page 2: ...adjusted to the ambient temperature be fore trying a new start up 2 General instructions and safety regulations for electric light and heavy current plants including domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third party devices e g connection GND and earth...

Page 3: ...6 1 Assignment of elbow type plug 4 20 mA 2 wire connection voltage 3 wire connection 1 supply 2 GND signal 1 signal 3 supply voltage Uv 4 supply voltage Uv The type current or voltage output is set by works and cannot be changed 6 2 Terminal assignment of meas cell socket contact no 4 pole cell K 0 55 1 temperature 2 temperature GND 3 supply 1 4 signal 1 5 signal 2 6 supply 2 7 not used 6 3 Usage...

Page 4: ... should be controlled or improved this is done by means of suitable reference solutions and the adjusting of the cell factor Attention Wrong handling of reference solution can make them useless very fast Measuring procedure Especially when measuring low conductivity Before immersion to the measuring solution rinse the measur ing cell with deionised water The measuring is speeded up considerably if...

Page 5: ...CLr After 10 seconds the currently measured values will be displayed again 9 Error and system messages Display Description Possible fault cause Remedy Err 1 measuring range ex ceeded Wrong signal Temperature above 140 C not allowed Choose right measuring range for con ductivity Err 2 Measuring value below measuring range Wrong signal Temperature below 5 C not allowed Choose right measuring range f...

Page 6: ...set above Possible inputs Range rAnG r 4 Range rAnG r 3 Range rAnG r 2 Range rAnG r 1 Range rAnG r 0 Conductivity cond 0 500 mS cm 0 0 200 0 mS cm 0 00 20 00 mS cm 0 2000 µS cm 0 0 200 0 µS cm Resistivity rESi 1 00 50 00 Ohm cm 1 0 500 0 Ohm cm 1 5000 Ohm cm 0 00 20 00 kOhm cm 0 200 kOhm cm TDS tdS 0 1000 g l 0 0 200 0 g l 0 00 20 00 g l 0 2000 mg l 0 0 200 0 mg l Salinity SAL 0 0 70 0 10 4 t Cor ...

Page 7: ...e measuring The offset of the measuring will be shifted by this value the input is in C Calculation see below Max input range 5 0 5 0 C or oFF 0 0 factory setting off 10 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input range 5 00 5 00 or oFF 0 00 factory setting off The adjusting of the temperature...

Page 8: ... mA type 2 wire output signal electrically isolated options AV01 0 1 V AV10 0 10 V 3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20 mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10 mA 0 10 V or refer to type plate Rev voltage protection 50 V permanent Perm impedance at 4 20mA RA Ohm Uv 12 V 0 02 A Example for Uv 18V RA 18 V 12 V 0 02 A RA 300 Ohm Per...

Reviews: