6 Configuration of the device
Follow these instructions to configure the functions of the device:
Switch the device off.
Press
and keep it pressed, while turning the device on (press
shortly).
Release the sort-
button not before the first parameter “
P.OF
” is displayed.
Set parameter with up
or down
.
Jump to the next parameter by pressing
.
Parameter Value
Information
button
buttons
.
Auto Power-Off
(turn-off delay)
factory setting: 20 min.
1
…
120
Auto Power-Off
(turn-off delay) in minutes.
If no key is pressed for the time adjusted here, the device is automati-
cally switched off (adjustable 1 .. 120 min.).
0F
Auto power-off is deactivated (continuous operation)
.
Display unit of moisture
factory setting: %u
%u
Measuring value displayed in moisture %u
%w
Measuring value displayed in water content %w
.
Display unit of temperature input
factory setting: °C
°
(
Temperature input in °C
°
F
Temperature input in °F
.
Auto Hold
factory setting: oF
0F
AutoHold deactivated: pressing hold-key freezes measuring and re-
leases it again.
ON
AutoHold activated:
hold-key starts new measuring, the display will
be ‘frozen’ as soon as a stable measuring value
was detected.
Press
again to store the changed settings, the device restarts (segment test).
NOTE:
If there is no key pressed within the menu mode within 2 minutes, the configuration will
be cancelled, the entered settings are lost!
7 Error and System Messages
E. 1
Value exceeding measuring range, value too high
E. 7
System error - the device has detected a system error
(device defective or not within operating temperature)
Value below display range
The blinking bAt display indicates low battery voltage, device will continue to work for a
short time.
The battery is consumed and has to be changed.
Measurements are no longer possible.