Tap lightly with the flat of your hand on the
Multi-MagPen Sleeve. This will mechanically detach
the 3D cell cultures from the Multi-MagPen Sleeve
in addition to the magnetic force of the Holding
Drive. Remove the Multi-MagPen Sleeve and cover
the receiver plate with its original lid.
Place the receiver plate containing
only
the
Multi-MagPen Sleeve on the 24 Well Holding Drive
(part of the 24 Well Bio-Assembler Kit; Item No.
662840). The 3D cell cultures will detach from the
Multi-MagPen Sleeve by the magnetic force of the
Holding Drive and settle on the bottom of the well.
WARNING
DO NOT place the receiver plate still containing the Multi-MagPen Drive in the
Multi-MagPen Sleeve on the Holding Drive. Due to the magnetic force they
can strongly attract each other, and separation of the components becomes
difficult and may damage the magnet drives.
Take out the Multi-MagPen Drive from the
Multi-MagPen Sleeve.