Instructions to transfer 3D cell cultures using the 24 Well Multi-MagPen

To assure sterility clean the Multi-MagPen Drive with 70 % ethanol. DO NOT autoclave the Multi-

 

MagPen 

Drive (see also safety precautions). During the cellular transfer process, the 

 

Multi-MegPen Drive is 

 

inserted into the sterile Multi-MagPen Sleeve. Hence the Multi-MagPen Drive is never in direct contact 

with the cell culture itself. 

If media must be removed from the 
wells, the microplate should be placed 
on the 24 Well Holding Drive (part of 
the 24 Well Bio-Assembler Kit; Item 
No. 662840) to secure the 3D cell 

 

cultures at the bottom of the wells 
during pipetting.

Prepare Receiver Plate

The receiver plate should contain the same liquid 
volume as the donor plate (1200 - 1500 µl/well). If 
the media volume is too low, the Multi-MagPen 
Sleeve will not touch or dip in the liquid and 3D cell 
cultures will not be transferred properly to the 
new plate. If the media volume is too high, media 
will spill out the well as soon as the Multi-MagPen 
Sleeve will dip in and 3D cell cultures might get lost.

Prepare Donor Plate

To assure liquid contact and reliable attachment 
of the 3D cell cultures to the Multi-MagPen Sleeve 

the recommended working volume per well is 

1200 – 1500 μl

. If the media volume is too low, the 

Multi-MagPen Sleeve will not touch or dip in the 
liquid and 3D cell cultures will not attach to it. If the 
media volume is too high, media will spill out the 
well as soon as the Multi-MagPen Sleeve will dip in. 

3/ INSTRUCTIONS

2

Summary of Contents for 657824

Page 1: ...INSTRUCTIONS FOR USE 24 WELL MULTI MAGPEN Item No 657824 Your smart assistant for 3D cell culture transfer...

Page 2: ...to mag nets take precautions so that objects do not slam together Neodymium magnets are britt le and can shatter or crack sometimes producing dangerous fragments moving at high speeds Fingers can also...

Page 3: ...use only and not approved for human or animal use MULTI MAGPEN KIT SLEEVE Item No Product Description Content Packaging 657824 24 Well Multi MagPen Kit 24 Well Multi MagPen Drive transparent non steri...

Page 4: ...l cultures at the bottom of the wells during pipetting Prepare Receiver Plate The receiver plate should contain the same liquid volume as the donor plate 1200 1500 l well If the media volume is too lo...

Page 5: ...ulti MagPen Drive into the Multi MagPen Sleeve 4 Carefully make 2 3 circular movements with the plate on the desk to facilitate 3D cell culture attachment to the Multi MagPen Sleeve 5 Gently lift the...

Page 6: ...Place the receiver plate containing only the Multi MagPen Sleeve on the 24 Well Holding Drive part of the 24 Well Bio Assembler Kit Item No 662840 The 3D cell cultures will detach from the Multi MagP...

Page 7: ...eeve Put the Multi MagPen Sleeve again into the receiver plate which is still located on the Holding Drive and repeat step 9 and 10 in order to improve the detachment of the 3D cell cultures Inconsist...

Page 8: ...GmbH has all copyrights and or other user rights in this documents in particular to signs such as the mentioned word picture brands and logos Any use duplication or any other use of the rights of Gre...

Reviews: