7.2 Limit of the Instrument
SRT 10/II reading limits is max. 140 mm. at 18°C. When a sample has a value > 140 mm,
this message is displayed or printed: “
> 140
”
8. ERROR INFORMATION AND WARNINGS
8.1 Warnings information
This message appears when printer is enabled in the SRT 10/II, but not switched on.
Meaning of symbol on display
----------------------------------------------
"xxx" Result = xxx
" . "
Awaiting sample
">xxx"
Sample value higher than "xxx"
"lev" Error: initial level error
"rem" Error: removed sample
8.2 System Error Warnings
This warning will be given if the instrument finds problems with the mechanical movement
of the reading plate during the initial self-test. (Call service)
After this indication the instrument will stop and it need to call the technical service.
This message can appear if there is a mechanical problem during analysis.
This message can appear if the internal thermometer has problems. In this case, analysis
results are displayed without temperature compensation.
18°C temp. of reference
temp. sensor error...
ERROR: call service...
ERROR: system stopped...
check the printer!