Typical Measurements and Accuracy

Electrical Accuracy Specifications

The Temperature Coefficient is 20% of accuracy per °C difference from 18 °C between 
0 °C and 18 °C and from 28 °C between 28 °C and 50 °C.

Accuracy

± (%reading)

Temperature

23

° 

C±5

° 

C

Humidity

≤ 80% RH

Temperature Coefficient

0.2 x (accuracy)/

°

C

(1) FC-110/118 AC Current Measurement

Range

Resolution

Accuracy (At 
Centered Position)

Cable Position

30.0A

0.1A

±(3%+0.5A)

Accuracy when 
measuring 
in centered 
position.

Please refer  
to figures 9  
and 10.

300A

1A

±(3%+5A)  

3000A

10A

±(3%+50A) 

Frequency Response

45Hz-500Hz

FC-118 Accuracy

Additional 
accuracy range 
when measuring 
outside of 
optimum location 
(when no other 
electrical or 
magnetic fields  
are present)

Center for optimal 
measurement

± (3%+5)

35mm (1.4 inches)

Add 1.0%

Region A

50mm (2.0 inches)

Add 1.5%

Region B

60mm (1.4 inches)

Add 2.0%

Region C

FC-110 Accuracy

Additional 
accuracy range 
when measuring 
outside of 
optimum location

(When no other 
electrical or 
magnetic fields 
are present)

Central optimum 
measurement 
location

±(3%+5)

FC-110

15mm 
(0.6”) away  
from center

Additional 2.0%

Region A

25mm  
(1.0”) away  
from center

Additional 2.5%

Region B

35mm  
(1.4”) away  
from center

Additional 3.0%

Region C

FC-110

Figure 9

Figure 10

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for FC-110

Page 1: ...idad que aparecen en este manual antes de manejar estas herramientas o darles mantenimiento Lire attentivement et bien comprendre toutes les instructions et les informations sur la s curit de ce manue...

Page 2: ...ex Clamp Meters are hand held testing devices that measure AC current These meters are designed to be placed on or removed from insulated or uninsulated conductors Safety Safety is essential in the us...

Page 3: ...injury or property damage Read and understand this material before operating or servicing this equipment Failure to understand how to safely operate this tool could result in an accident causing seri...

Page 4: ...e injury or death Electric shock hazard Do not apply more than the rated voltage between any two input terminals or between any input terminal and earth ground Keep hands and fingers below the barrier...

Page 5: ...flash when the batteries are low 5 Range Switch Select the appropriate current range for the application Ranges are 30A 300A and 3000A 6 Output Voltage Description Lists the output voltage resolution...

Page 6: ...g uninsulated conductors do not power on the wire to be tested until the meter has safely clamped the wire 1 Turn off the clamp meter as well as the conductor being measured 2 Unlock the clamp accordi...

Page 7: ...greater than 3V will display on the AC voltage measuring device Please select the appropriate range 30 00A 300 0A 3000A 7 Improper operation examples a Do not test more than one conductor at a time as...

Page 8: ...0A Frequency Response 45Hz 500Hz FC 118 Accuracy Additional accuracy range when measuring outside of optimum location when no other electrical or magnetic fields are present Center for optimal measure...

Page 9: ...Measurement Categories These definitions were derived from the international safety standard for insula tion coordination as it applies to measurement control and laboratory equipment These measureme...

Page 10: ...vere injury or death a The repair and service of the meter should be accomplished by Greenlee professional maintenance personnel or authorized departments b Clean the meter case by using dry cloth per...

Reviews: