CMP-200
21
* d refers to the least significant digit (the number in the right-most position)
** Sensitivity: Volts > 1 V; Amps > 5 A
Minimum Input Frequency: 10 Hz
† Maximums: 600 V, 2000 Amps
Power Factor: 0.2 to 1.0
†† Accuracy specified for crest factor less than 4
Accuracy
Accuracy is specified as follows:
±
(a percentage of the r a fixed amount) at 18
°
C to
28
°
C (64.4
°
F to 82.4
°
F).
Accuracy Table
Value
Measurement Range
Accuracy
**Frequency
50/60 Hz
±
(2 d)*
10 Hz to 1000 Hz
±
(1.5% + 2 d)*
0 to 99.99 KW
±
(2% + 0.05 KW)
†True Power (AC or DC)
100 to 999.9 KW
±
(2% + 0.5 KW)
1000 to 1200 KW
±
(2% + 5 KW)
0 to 99.99 KW
±
(2% + 0.05 KW)
†KVAR (AC or DC)
100 to 999.9 KW
±
(2% + 0.5 KW)
1000 to 1200 KW
±
(2% + 5 KW)
at O Hz and 50/60 Hz
at 40 Hz to 400 Hz
††Voltage (AC or DC) at
0 to 499.9 V
±
(1.5% + 0.5 V)
±
(2.0% + 0.5 V)
10 m
Ω
Input Impedance
500 to 600 V
±
(1.5% + 5 V)
±
(2.0% + 5 V)
0 to 199.9 A
±
(1.5% + 0.5 A)
±
(2.0% + 0.5 A)
††Current (AC or DC)
200 to 499.9 A
±
(2.0% + 0.5 A)
±
(2.5% + 0.5 A)
500 to 2000 A
±
(2.0% + 5 A)
±
(2.5% + 5 A)