Greenlee CLM-1000 Instruction Manual Download Page 6

6

Operation

Electric shock hazard:
Do not connect the unit to live voltage.
Contact with live circuits could result 

in severe injury or death.

Calibration Procedure

Each time the meter is turned ON, it should be calibrated 

before use.
1.  Refer to “Typical Measurements” for calibration setup 

diagram.

2.  Ensure the calibration bar and Kelvin clips jaws are 

clean.

3.  Move rotary switch from the 

OFF

 position to 

Ω

.

4.  Ensure that both leads for the red set of Kelvin clips 

are plugged into the red terminals on the meter.

 

Ensure that both leads for the black set of Kelvin clips 

are pluggged into the black terminals on the meter.

5.  Place each Kelvin clip on the calibration bar. Move the 

clips as close together as possible.

6.  Press the 

CAL

 button. Dashes appear on the LCD. 

Next “MEM” and “donE” appear on the LCD. Then 

“RES” on the LCD indicates the meter is back in 

resistance mode. If “FAIL” appears on the LCD and the 

buzzer sounds, there was an issue with the calibration 

process. Turn meter OFF and return to step 1 of this 

procedure.

7.  After successful completion of the calibration process, 

the meter is ready for normal operation.

Operating Procedure

1.  If the unit does not function as expected on a known 

functioning configuration, replace the batteries.

 

If the unit still does not function as expected, call 

Greenlee for technical assistance at 800-435-0786.

2.  Refer to “Typical Measurements” for specific mea-

surement instructions.

3.  The meter does not compensate for manufacturing 

tolerances of various cable manufacturers.

4.  Before making a measurement, ensure the following:

• Cable conductors have been cleaned with 

abrasive material to remove oxide layer and other 

contaminants.

• Kelvin clip jaws are clean.
• Both the cable and meter are at the same 

temperature.

• Cable conductor material must be either aluminum 

or copper.

5.  To measure odd cable sizes:

• 

AWG (5, 7, 9, 11 AWG, etc.) (CLM-1000): 

When 

measuring odd numbered cable sizes, set the rotary 

switch to the next smaller size, and multiply the 

reading by 1.261.
For example, to measure a 9 AWG length of cable, 

set the rotary switch to 

10 AWG

. Then multiply the 

reading by 1.261 to get the length.

• 

kcmil (CLM-1000): 

When measuring odd 

numbered cable sizes, set the rotary switch to 

1000 kcmil

. Divide kcmil size of cable being mea-

sured by 1000, and then multiply this number by the 

meter’s length reading.
For example, 477 kcmil cable is being measured. 

The multiplier becomes 477/1000 = 0.477. Take 

the value shown with the rotary switch in the 1000 

kcmil position and multiply it by 0.477 to get the 

length.

• 

mm

2

 (CLM-1000E): 

When measuring odd 

numbered cable sizes, set the rotary switch to 

500 mm

2

. Divide mm

2

 size of cable being mea-

sured by 500, and then multiply this number by the 

meter’s length reading.
For example, 289 mm

2

 cable is being measured. 

The multiplier becomes 289/500 = 0.596. Take the 

value shown with the rotary switch in the 500 mm

2

 

position and multiply it by 0.596 to get the length.

Programming Procedure

1.  Calibrate meter according to “Calibration Procedure.”
2.  Obtain known length of cable between 30 ft and 

320 ft (9 m and 97.5 m) long. Cable must be alumi-

num or copper.

3.  Set rotary switch to desired location. Programmable 

location appears on the LCD. If the location has 

already been programmed, “OL” appears on the LCD. 

If it has not, “no” appears.

Note:  Data in a programmed location can be deleted. 

With Kelvin clips removed from cable and rotary 

switch turned to location to delete, press 

MEM

 button.

4. Press 

CAL

 button.

5.  Enter the known length of cable in feet or meters. 

Use 

qp

 arrows.

6.  Select “Cu” or “Al”.
7.  Connect Kelvin clips to the conductor on each end of 

the cable.

8. Press 

MEM

 button. “MEM” and dashes appear on the 

LCD. After approximately 1 second, “donE” appears 

on the LCD if programming was successful. If “FAIL” 

appears, check connections and repeat procedure.

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for CLM-1000

Page 1: ...toutes les instructions et les informations sur la s curit de ce manuel avant d utiliser ou de proc der l entretien de cet outil Vor Bedienung oder Wartung dieses Messger tes bitte alle Anweisungen u...

Page 2: ...150 185 240 mm2 61 strands 300 400 500 Other capabilities include Automatic temperature compensation Hold Mode to capture the present LCD value Programmable Auto Power Off Automatic ranging Backlight...

Page 3: ...connect this unit to live voltage Do not expose this unit to rain or moisture Do not use the unit if it is wet or damaged Using this unit near equipment that generates electromagnetic interference ca...

Page 4: ...struction manual Warning Risk of electric shock Double insulation Display Icons 1 Low battery indicator 2 User selected switch position 3 Ambient temperature indication 4 Resistance Mode 5 Copper 6 Al...

Page 5: ...e press the HOLD FT M or Cu Al button dH is no longer present on the LCD Auto Power Off APO When enabled the meter will automatically turn OFF after 10 minutes of inactiv ity Inactivity is defined as...

Page 6: ...e odd cable sizes AWG 5 7 9 11 AWG etc CLM 1000 When measuring odd numbered cable sizes set the rotary switch to the next smaller size and multiply the reading by 1 261 For example to measure a 9 AWG...

Page 7: ...Typical Measurements Typical Configuration Calibration Note Make sure that the clips are as close together as possible in order to improve calibration accuracy Calibration Bar www com information itm...

Page 8: ...0 01 kft 100 00 kft 1 5 0 33 kft Length in meters Gauges CLM 1000 Diameter mm2 CLM 1000E Accuracy Range Length Range Accuracy 18 AWG 2 AWG 0 75 35 1000 0 m 5 m 152 4 m 1 0 1 6 m 152 5 m 304 8 m 1 0 2...

Page 9: ...h ISO 9001 2000 for our Quality Management Systems The instrument enclosed has been checked and or cali brated using equipment that is traceable to the National Institute for Standards and Technology...

Reviews: