Measurements
(cont’d)
Fig. 13
Set-Up Test Voltage in Insulation
Resistance Measurement
Fig. 13a
Set-Up Insulation
Resistance Limit Value
Set-up test voltage
for Insulation Resistance (
Fig.13)
:
Adjust test voltage using the
t
and
u
buttons.
Legend of displayed symbols:
Name of selected function
Insulation Resistance
Selected test voltage
Nominal
250V
Set-Up Insulation Resistance Limit Value
for
Insulation Resistance
(Fig.13):
• Press the Ω button
• Adjust the limit value using the
t
and
u
buttons
• Press the Ω button again or the
START
button to
return to the insulation resistance measurement menu.
Legend of displayed symbols:
Name of selected function
Insulation Resistance
Selected Insulation Resistance Limit Value
Resistance Value 1 MΩ
Insulation resistance limit selection range:
[10 kΩ .. 2--MΩ]
No limit selected indication:
– – –
Warning!
Refer to Warnings chapter for safety precautions!