3 Operation
30
Series 330 Vacuum Gauge Controller
Instruction Manual - 330004 - Rev. 02
3.3 Specifications
3.3.2 Electrometer/Ion Gauge Pressure Range Specifications
An internal overpressure limiter is factory adjusted to trip at a pressure of 1 decade below the upper
limits specified above. Refer to Section 4.4.3 for readjustment instructions.
3.3.1 330 VGC Specifications
Table 3-2 Series 330 VGC Physical Specifications
Parameter
Specification
Width
241 mm (9.5 in.) with 1/2 rack mounting brackets
Height
89 mm (3.5 in.)
Depth
356 mm (14 in.) includes 76 mm (3 in.) for connectors and cables
Weight
4.8 kg (10.5 lb.)
Environmental Temperature
0
o
C to 40
o
C
Table 3-3 Series 330 VGC Electrical Specifications
Parameter
Specification
Voltage
90-130 VAC or 200-260 VAC
Frequency
50 to 60 Hz
Power
100 watts max
Fuse Ratings
1.25 A (90-130 V)
.60 A (200-260 V)
Table 3-4 Emission Range for Non UHV Ion Gauges
.01mA to.1mA
.1mA to 1mA
1mA to 10mA
1 x 10
-8
to 1 x 10
-1
Torr
1 x 10
-9
to 1 x 10
-2
Torr
1 x 10
-10
to 1 x 10
-3
Torr
Readable to:
1 x 10
-9
Torr
1 x 10
-10
Torr
1 x 10
-11
Torr
Table 3-5 Pressure Display Specifications
Parameter
Specification
Electronic Accuracy
Typical ± 3% of reading at ambient temperature of 25 ± 5
o
C
Display Units
Torr unless otherwise requested. Adjustment and internal selector switch provides readout
in mbar or pascal.
Display Resolution
Scientific notation, 2 significant digits
Display Update Time
0.5 sec. typical as shipped. Internal switch selectable to 3 sec. reading averaged
Summary of Contents for Series 330
Page 2: ......
Page 8: ...Table of Contents 8 Series 330 Vacuum Gauge Controller Instruction Manual 330004 Rev 02...
Page 75: ......