ecocool 150R
Operating Manual
Page 36
www.grantinstruments.com
1. Press the
F
and
S
buttons simultaneously for 3
seconds.
The calibrate probe menu displays the factory
defined calibration points and the current
temperature of the bath in the top right corner of
the screen
.
2. Rotate the dial to highlight
add point
, press
S
to
select.
The calibrate point menu is displayed. If an
external probe has been selected the external
probe icon will be displayed in the top left corner.
Otherwise the settings icon will be displayed when
calibrating the internal probe.
3. Rotate the dial to highlight
measured
temperature
. Press
S
to select
4. Rotate the dial until the temperature being shown
on the calibrated thermometer is displayed on
screen. Press
S
to save the value.
5. Rotate the dial to highlight
save
, press
S
to select.
The display returns to the calibrate probe screen.
6. Press
F
to return to the home screen
Further calibration points can be added by
repeating steps 2 to 5.
Calibration points can be deleted by selecting
delete in the calibrate point screen.
.
°c
Point 19.86 c
Calibrate Probe
Point 70.04 c
Add Point
Restore Factory
.
°c
Measured Temperature
Calibrate Point
°c
.
Save
Delete